Abstract
A culture-based colony counting method is the most widely used analytical technique for monitoring bioaerosols in both indoor and outdoor environments. However, this method requires several days for colony formation. In this study, our goal was fast monitoring (Sampling: 3 min, Detection: < 1 min) of indoor bioaerosol concentrations with ATP bioluminescence assay using a bioaerosol sampler. For this purpose, a novel hand-held electrostatic rod-type sampler (110 mm wide, 115 mm long, and 200 mm tall) was developed and used with a commercial luminometer, which employs the Adenosine triphosphate (ATP) bioluminescence method. The sampler consisted of a wire-rod type charger and a cylindrical collector, and was operated with an applied voltage of 4.5 kV and a sampling flow rate of 150.7 lpm. Its performance was tested using Staphylococcus epidermidis which was aerosolized with an atomizer. Bioaerosol concentrations were measured using ATP bioluminescence method with our sampler and compared with the culture-based method using Andersen cascade impactor under controlled laboratory conditions. Indoor bioaerosol concentrations were also measured using both methods in various indoor environments. A linear correlation was obtained between both methods in lab-tests and field-tests. Our proposed sampler with ATP bioluminescence method may be effective for fast monitoring of indoor bioaerosol concentrations.Keywords
This publication has 36 references indexed in Scilit:
- Development of an Automated Electrostatic Sampler (AES) for Bioaerosol DetectionAerosol Science and Technology, 2011
- Collection efficiencies of an electrostatic sampler with superhydrophobic surface for fungal bioaerosolsIndoor Air, 2010
- A Handheld Electrostatic Precipitator for Sampling Airborne Particles and NanoparticlesAerosol Science and Technology, 2010
- Performance of an Electrostatic Precipitator with Superhydrophobic Surface when Collecting Airborne BacteriaAerosol Science and Technology, 2010
- Design and Application of an Inertial Impactor in Combination with an ATP Bioluminescence Detector for In Situ Rapid Estimation of the Efficacies of Air Controlling Devices on Removal of BioaerosolsEnvironmental Science & Technology, 2010
- Application of ATP bioluminescence method to characterize performance of bioaerosol sampling devicesJournal of Aerosol Science, 2009
- Size distributions of total airborne particles and bioaerosols in a municipal composting facilityBioresource Technology, 2008
- Field detection and identification of a bioaerosol suite by pyrolysis‐gas chromatography‐ion mobility spectrometry*Field Analytical Chemistry & Technology, 2001
- LIVE/DEAD® BacLight™: application of a new rapid staining method for direct enumeration of viable and total bacteria in drinking waterJournal of Microbiological Methods, 1999
- Diurnal Distribution of Total and Culturable Atmospheric Bacteria at a Rural SiteAerosol Science and Technology, 1999