Effect of the Coronal Wall Thickness of Dental Implants on the Screw Joint Stability in the Internal Implant-Abutment Connection

Authors:
  • JLJi-Hye Lee
    Ji-Hye Lee
  • YHYoon-Hyuk Huh
    Yoon-Hyuk Huh
  • CPChan-Jin Park
    Chan-Jin Park
  • LCLee-Ra Cho
    Lee-Ra Cho

Abstract
Ji-Hye Lee, DDS, PhD/Yoon-Hyuk Huh, DDS, MSD/Chan-Jin Park, DDS, MSD, PhD/Lee-Ra Cho, DDS, MSD, PhD: Purpose: To evaluate the effect of implant coronal wall thickness on load-bearing capacity and screw joint stability. Materials and Methods: Experimental implants were customized after investigation of the thinnest coronal wall thickness of commercially available implant systems with a regular platform diameter. Implants with four coronal wall thicknesses (0.2, 0.3, 0.4, and 0.5 mm) were fabricated. Three sets of tests were performed. The first set was a failure test to evaluate load-bearing capacity and elastic limit. The second and third sets were cyclic and static loading tests. After abutment screw tightening of each implant, vertical cyclic loading of 250 N or static loading from 250 to 800 N was applied. Coronal diameter expansion, axial displacement, and removal torque values of the implants were compared. Repeated measures analysis of variance (ANOVA) was used for statistical analysis (α = .05). Results: Implants with 0.2-mm coronal wall thickness demonstrated significantly low load-bearing capacity and elastic limit (both P < .05). These implants also showed significantly large coronal diameter expansion and axial displacement after screw tightening (both P < .05). Greater vertical load and thinner coronal wall thickness significantly increased coronal diameter expansion of the implant, axial displacement of the abutment, and removal torque loss of the abutment screw (all P < .05). Conclusion: Implant coronal wall thickness of 0.2 mm produces significantly inferior load-bearing capacity and screw joint stability.