Effect of Improper Scan Alignment on Retinal Nerve Fiber Layer Thickness Measurements Using Stratus Optical Coherence Tomograph
- 1 August 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 17 (5) , 341-349
Abstract
Objective Misalignment of the Stratus optical coherence tomograph scan circle placed by the operator around the optic nerve head (ONH) during each retinal nerve fiber layer (RNFL) examination can affect the instrument reproducibility and its theoretical ability to detect true structural changes in the RNFL thickness over time. We evaluated the effect of scan circle placement on RNFL measurements. Design Observational clinical study. Methods Sixteen eyes of 8 normal participants were examined using the Stratus optical coherence tomograph Fast RNFL thickness acquisition protocol (software version 4.0.7; Carl Zeiss Meditec, Dublin, CA). Four consecutive images were taken by the same operator with the circular scan centered on the optic nerve head. Four images each with the scan displaced superiorly, inferiorly, temporally, and nasally were also acquired. Differences in average and sectoral RNFL thicknesses were determined. For the centered scans, the coefficients of variation (CV) and the intraclass correlation coefficient for the average RNFL thickness measured were calculated. Results When the average RNFL thickness of the centered scans was compared with the average RNFL thickness of the displaced scans individually using analysis of variance with post-hoc analysis, no difference was found between the average RNFL thickness of the nasally (105.2 μm), superiorly (106.2 μm), or inferiorly (104.1 μm) displaced scans and the centered scans (106.4 μm). However, a significant difference (analysis of variance with Dunnett's test: F=8.82, P<0.0001) was found between temporally displaced scans (115.8 μm) and centered scans. Significant differences in sectoral RNFL thickness measurements were found between centered and each displaced scan. The coefficient of variation for average RNFL thickness was 1.75% and intraclass correlation coefficient was 0.95. Conclusions In normal eyes, average RNFL thickness measurements are robust and similar with significant superior, inferior, and nasal scan displacement, but average RNFL thickness is greater when scans are displaced temporally. Parapapillary scan misalignment produces significant changes in RNFL assessment characterized by an increase in measured RNFL thickness in the quadrant in which the scan is closer to the disc, and a significant decrease in RNFL thickness in the quadrant in which the scan is displaced further from the optic disc.Keywords
This publication has 20 references indexed in Scilit:
- Optical coherence tomography for clinical detection and monitoring of glaucoma?British Journal of Ophthalmology, 2007
- Combining Structural and Functional Testing for Detection of GlaucomaOphthalmology, 2006
- Influence of Pupil Size and Cataract on Retinal Nerve Fiber Layer Thickness Measurements by Stratus OCTJournal of Glaucoma, 2006
- Structure–Function Relationships Using Confocal Scanning Laser Ophthalmoscopy, Optical Coherence Tomography, and Scanning Laser PolarimetryInvestigative Ophthalmology & Visual Science, 2006
- Diagnostic Capability of Optical Coherence Tomography in Evaluating the Degree of Glaucomatous Retinal Nerve Fiber DamageInvestigative Ophthalmology & Visual Science, 2006
- Relationship Between Patterns of Visual Field Loss and Retinal Nerve Fiber Layer Thickness MeasurementsAmerican Journal of Ophthalmology, 2006
- Diagnostic Ability of Optical Coherence Tomography with a Normative Database to Detect Localized Retinal Nerve Fiber Layer DefectsOphthalmology, 2005
- Sensitivity and specificity of the StratusOCT for perimetric glaucomaOphthalmology, 2005
- Quantitative assessment of structural damage in eyes with localized visual field abnormalitiesAmerican Journal of Ophthalmology, 2004
- Thickness of the peripapillary retina in healthy subjects with different degrees of ametropiaOphthalmology, 2000