Effect of Repetitive Loading on the Mechanical Properties of Synthetic Hernia Repair Materials
Authors:
- Corey R. DeekenCDCorey R. Deeken
- 1 September 2011
- journal article
- Published by Wolters Kluwer Health in Journal of the American College of Surgeons
- Vol. 213 (3) , 430-435
Abstract
Hernia repair materials undergo repeated loading while in the body, and the impact on mechanical properties is unknown. It was hypothesized that exposure to repetitive loading would lead to decreased tensile strength and increased strain, and that these differences would become more pronounced with greater loading and unloading sequences. Polypropylene, expanded polytetrafluoroethylene, composite barrier, and partially absorbable meshes were evaluated. Twenty specimens (7.5 × 7.5 cm) were prepared from each material. Five specimens were subjected to ball burst testing to determine baseline biomechanical properties. Cycles of 10, 100, and 1,000 loading sequences were also performed (n = 5 each). BardMesh (CR Bard/Davol), Dualmesh (WL Gore), and Prolene (Ethicon) exhibited significantly reduced tensile strength; BardMesh, Proceed (Ethicon), Prolene, ProLite (Atrium Medical), ProLite Ultra (Atrium Medical), and Ultrapro (Ethicon) exhibited significantly increased strain after exposure to 1,000 cycles compared with their baseline properties. BardMesh and Prolene demonstrated both reduced tensile strength and increased strain values after 1,000 cycles, suggesting that repetitive loading has the greatest effects on these materials. In addition, BardMesh and Prolene exhibited progressively worsening effects as the number of cycles was increased. Deterioration of the tensile strength of the mesh or an increase in the ability of the mesh material to stretch (ie, increased strain values) could potentially lead to hernia recurrence or a poor functional result. However, the results of this study should not be interpreted to mean that hernia repair materials will fail in the body after only 10, 100, or 1,000 cycles. The conditions used in this study were more extreme than most physiologic scenarios and were intended as a pilot investigation into how the mechanical properties of hernia repair materials are affected by in vitro cyclic testing.Keywords
This publication has 9 references indexed in Scilit:
- Physicomechanical Evaluation of Polypropylene, Polyester, and Polytetrafluoroethylene Meshes for Inguinal Hernia RepairJournal of the American College of Surgeons, 2011
- Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repairSurgical Endoscopy, 2010
- The extracellular matrix as a biologic scaffold materialBiomaterials, 2007
- Normal Intraabdominal Pressure in Healthy AdultsJournal of Surgical Research, 2005
- The lightweight and large porous mesh concept for hernia repairExpert Review of Medical Devices, 2005
- The Use of a Subfascial Vicryl Mesh Buttress to Aid in the Closure of Massive Ventral Hernias following Damage-Control LaparotomyPlastic and Reconstructive Surgery, 2003
- Elasticity of the anterior abdominal wall and impact for reparation of incisional hernias using mesh implantsHernia, 2001
- Functional assessment and tissue response of short- and long-term absorbable surgical meshesBiomaterials, 2001
- Use of an Absorbable Mesh to Repair Contaminated Abdominal-Wall DefectsArchives of Surgery, 1986