The significance of implant shape in experimental testing of biological materials: Disc vs. Rod
Authors:
- 1 March 1970
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 4 (1) , 1-12
Abstract
Adult albino rabbits were implanted with 304 stainless steel discs and cylinders (rods) in the following sites: beneath periosteum of the anterior calvarium, under the masseter muscle of the left mandibular ramus and in the body of the sacrospinalis muscle. (1) The tissue reaction around the disc implants was compared with that observed around the cylinder implants through the study of histological sections. (2) The degree of reactions observed at the various sites was contrasted with that seen at the other sites. In all cases the muscle implants showed the greatest reaction. Discs showed many micro areas of tissue reaction randomly around their periphery. In contrast all the rod shaped implants in muscle showed a greater reaction towards the ends than in the mid portion of the shaft (clubbing) but rods implanted at other sites did not show this phenomenon. The results show that much of the histological tissue reaction seen around muscle implants is really caused by mechanical trauma which must be differentiated from a non‐compatibility reaction. It is important to choose a test site where mechanical trauma will be minimal (submasseteric site) and use an implant shape (discs) which will not produce clubbing.Keywords
This publication has 12 references indexed in Scilit:
- Toxicity of Metal Particles in Tissue CultureJournal of Bone and Joint Surgery, 1968
- The behavior of biological materials at different sites of implantationJournal of Biomedical Materials Research, 1968
- Tissue reaction in rabbit muscle exposed to metallic implantsJournal of Biomedical Materials Research, 1967
- Characteristics of Trace Ions Released from Embedded Metal Implants in the RabbitJournal of Bone and Joint Surgery, 1962
- Corrosion Testing of Orthopaedic ImplantsJournal of Bone and Joint Surgery, 1962
- Tissue Reactions to Metals—The Influence of Surface FinishJournal of Bone and Joint Surgery, 1961
- Experimental Production of Cancer by Means of Implanted Polyurethane PlasticAmerican Journal of Clinical Pathology, 1960
- ELECTROLYSIS AND STAINLESS STEELS IN BONEThe Journal of Bone and Joint Surgery. British volume, 1956
- Bone Fixation and the Corrosion Resistance of Stainless Steels to the Fluids of the Human BodyJournal of the Electrochemical Society, 1948
- THE EFFECTS ON BONE OF THE PRESENCE OF METALS; BASED UPON ELECTROLYSISAnnals of Surgery, 1937