JET divertor diagnostic upgrade for neutral gas analysis
- 21 August 2012
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (10) , 10D728
Abstract
With installation of the ITER-like wall in JET a major diagnostic upgrade to measure the neutral gas pressure and composition in the sub-divertor region has been completed, to characterise retention and outgassing of the new metallic first wall. The upgrade includes two new magnetically shielded systems consisting of sensitive capacitance manometers and residual gas analysers, both capable of providing data during plasma operation. These enable absolute pressure and gas composition measurements (pressure range: 10−5–10−1 mbar, mass range: 1–200 amu, respectively) and have been used to characterise the neutral gas behaviour under various plasma conditions.Keywords
This publication has 4 references indexed in Scilit:
- JET ITER-like wall—overview and experimental programmePhysica Scripta, 2011
- Overview of experimental preparation for the ITER-Like Wall at JETJournal of Nuclear Materials, 2011
- Helium and neon enrichment studies in the JET Mark IIAP and Mark IIGB divertorsNuclear Fusion, 2002
- Tritium concentration measurements in the Joint European Torus divertor by optical spectroscopy of a Penning dischargeReview of Scientific Instruments, 1999