A new amplification strategy (ADRO™), based on 64 independently operating channels, was compared with a nine-channel wide dynamic range compression strategy (WDRC). Open-platform in-the-ear hearing instruments were configured either with ADRO or the manufacturer's WDRC strategy. Twenty-two subjects with mild to moderate hearing loss took home the ADRO or WDRC hearing aids. After three weeks' acclimatization, the aids were evaluated using monosyllables in quiet at 50 to 65 dB SPL and sentences in eight-talker babble. The acclimatization and evaluation were repeated in the second phase of the balanced reverse-block blind experimental design. The ADRO program showed a statistically significant mean advantage of 7.85% word score (95% confidence interval 3.19% to 12.51%; p = 0.002) and 6.41% phoneme score for the monosyllables in quiet (95% confidence interval 2.03% to 10.79%; p = 0.006). A statistically significant advantage of 7.25% was also found for the ADRO program in background noise (95% confidence interval 1.95% to 12.55%; p = 0.010). The results are consistent with earlier data for listeners with moderate to severe hearing loss. Spanish Se comparó una nueva estrategia de amplificación (ADRO™), basada en 64 canales operativos independientes, con la estrategia de compresión de rango dinámico amplio (WDRC). Se configuraron instrumento auditivos intraauriculares de plataforma abierta, tanto con ADRO como con la estrategia WDRC del fabricante. Se adaptaron auxiliares auditivos con ADRO o con WDRC a veintidós sujetos con pérdidas auditivas leves a moderadas. Luego de un período de acostumbramiento de tres semanas, los audífonos fueron evaluados utilizando monosilábicos en silencio entre 50 y 65 dB SPL, y con frases en medio del balbuceo de ocho hablantes. El acostumbramiento y la evaluación se repitieron en la segunda fase de este diseño experimental de bloque-reverso, balanceado y ciego. El programa ADRO mostró una ventaja media estadísticamente significativa de 7.85% en el puntaje de palabras (intervalo de confianza del 95% de 3.19% a 12.51%; p = 0.002) y 6.41% en el puntaje de fonemas para los monosilábicos en silencio (intervalo de confianza del 95% de 2.03% a 10.79%; p = 0.006). Se encontró una ventaja estadísticamente significativa de 7.25% para el programa ADRO en medio de ruido ambiental de fondo (intervalo de confianza del 95% de 1.95% a 12.55%; p = 0.010). Los resultados son consistentes con datos previos en sujetos con hipoacusias moderadas a severas.