Geometry Calibration of a Modular Stereo Cone-Beam X-ray CT System
Open Access
- 13 March 2021
- journal article
- research article
- Published by MDPI AG in Journal of Imaging
- Vol. 7 (3) , 54
Abstract
Compared to single source systems, stereo X-ray CT systems allow acquiring projection data within a reduced amount of time, for an extended field-of-view, or for dual X-ray energies. To exploit the benefit of a dual X-ray system, its acquisition geometry needs to be calibrated. Unfortunately, in modular stereo X-ray CT setups, geometry misalignment occurs each time the setup is changed, which calls for an efficient calibration procedure. Although many studies have been dealing with geometry calibration of an X-ray CT system, little research targets the calibration of a dual cone-beam X-ray CT system. In this work, we present a phantom-based calibration procedure to accurately estimate the geometry of a stereo cone-beam X-ray CT system. With simulated as well as real experiments, it is shown that the calibration procedure can be used to accurately estimate the geometry of a modular stereo X-ray CT system thereby reducing the misalignment artifacts in the reconstruction volumes.Keywords
Funding Information
- university research fund (UAntwerp BOF-GOA 2016 33927)
- Research Foundation - Flanders (FWO) (S004217N)
- European Commission through the INTERREG (0386)
This publication has 24 references indexed in Scilit:
- Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installationRadiation Oncology, 2016
- Fast and flexible X-ray tomography using the ASTRA toolboxOptics Express, 2016
- Deep Residual Learning for Image Recognition2016
- The ASTRA Toolbox: A platform for advanced algorithm development in electron tomographyUltramicroscopy, 2015
- A robust geometry estimation method for spiral, sequential and circular cone‐beam micro‐CTMedical Physics, 2012
- Automatic alignment and reconstruction of images for soft X-ray tomographyJournal of Structural Biology, 2012
- Reliable automatic alignment of tomographic projection data by passive auto‐focusMedical Physics, 2011
- Estimation of CT cone-beam geometry using a novel method insensitive to phantom fabrication inaccuracy: Implications for isocenter localization accuracyMedical Physics, 2011
- Direct determination of geometric alignment parameters for cone-beam scannersPhysics in Medicine & Biology, 2009
- Accurate technique for complete geometric calibration of cone‐beam computed tomography systemsMedical Physics, 2005