X‐ray depth‐dose characteristics of the Toshiba LMR‐16

Authors:
  • JMJoseph Mantel
    Joseph Mantel
  • HPHarold Perry
    Harold Perry
  • JWJames J. Weinkam
    James J. Weinkam

Abstract
The depth-dose characteristics of the Toshiba LMR-16 linear accelerator for 14-MeV x rays have been measured at an SSD of 100 cm using diodes and ion chambers. The surface dose and build-up depth both exhibit a considerable variation with field size. A new central axis model has been developed which takes account of these variations, and the agreement between the measured and computed data using this model is found to be excellent. Formulas are also presented to estimate the surface dose, buildup depth, and output factor as a function of field size.