Comparison of radiation dose to patient and staff for two interventional cardiology units: a phantom study
- 10 June 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 131 (3) , 399-403
Abstract
The purpose of this investigation was to measure the patient and staff dose during routine interventional cardiology procedures for an image intensifier-based and a flat detector system using a water phantom. The Integris BH3000 image intensifier-based (Philips) and the Axiom Artis flat detector-based (Siemens) angiography units were used in this study. The accuracy of tubes potential and irradiation timers and also internal dosimeters were verified and confirmed. A water phantom with a thickness of 18 cm was used for patient and staff dose measurements. For the Philips system, phantom entrance dose rates were 2.77 and 38.97 µGym2 s−1 during fluoroscopy and cineangiography. The respective dose rates for the Siemens were 1.98 and 13.46 µGym2 s−1. Phantom entrance dose rate was 28.5 and 65% higher for the Philips system during fluoroscopy and cineangiography, respectively. Comparing the scattered dose rates at the operator location showed that the flat detector-based Siemens system delivers five times lower dose to the operator in comparison with the image intensifier-based Philips unit. The results suggest that the decrease in received dose of the patient and staff is achievable using the flat detector system. In addition, application of lead curtain and glass is recommended to lower the cardiologist dose especially for the image intensifier-based Philips system.Keywords
This publication has 11 references indexed in Scilit:
- Do flat detector cardiac X-ray systems convey advantages over image-intensifier-based systems? Study comparing X-ray dose and image qualityEuropean Radiology, 2006
- Variations of Patient Doses in Interventional Examinations at Different Angiographic UnitsCardioVascular and Interventional Radiology, 2006
- Influence of patient thickness and operation modes on occupational and patient radiation doses in interventional cardiologyRadiation Protection Dosimetry, 2006
- Does Digital Acquisition Reduce Patients' Skin Dose in Cardiac Interventional Procedures? An Experimental StudyAmerican Journal of Roentgenology, 2004
- Comparison of a conventional and a flat-panel digital system in interventional cardiology proceduresBritish Journal of Radiology, 2004
- 100 years of observation on British radiologists: mortality from cancer and other causes 1897–1997British Journal of Radiology, 2001
- Radiation Protection of Angiographers in Interventional RadiologyJapanese Journal of Radiological Technology, 2001
- Influence of Radiographic Technique and Equipment on Absorbed Ovarian Dose Associated with Uterine Artery EmbolizationJournal of Vascular and Interventional Radiology, 2000
- Patient dose in radiologically guided interventional vascular procedures: conventional versus digital systems.Radiology, 1997
- The impact of digital imaging on patient doses during barium studiesBritish Journal of Radiology, 1995