Clinical efficiency of operating room-based sliding gantry CT as compared to mobile cone-beam CT-based navigated pedicle screw placement in 853 patients and 6733 screws
Open Access
- 14 September 2021
- journal article
- research article
- Published by Springer Nature in European Spine Journal
- Vol. 30 (12) , 3720-3730
Abstract
Purpose: Multiple solutions for navigation-guided pedicle screw placement are available. However, the efficiency with regard to clinical and resource implications has not yet been analyzed. The present study’s aim was to analyze whether an operating room sliding gantry CT (ORCT)-based approach for spinal instrumentation is more efficient than a mobile cone-beam CT (CBCT)-based approach. Methods: This cohort study included a random sample of 853 patients who underwent spinal instrumentation using ORCT-based or CBCT-based pedicle screw placement due to tumor, degenerative, trauma, infection, or deformity disorders between November 2015 and January 2020. Results: More screws had to be revised intraoperatively in the CBCT group due to insufficient placement (ORCT: 98, 2.8% vs. CBCT: 128, 4.0%; p = 0.0081). The mean time of patients inside the OR (Interval 5 Entry–Exit) was significantly shorter for the ORCT group (ORCT: mean, [95% CI] 256.0, [247.8, 264.3] min, CBCT: 283.0, [274.4, 291.5] min; p < 0.0001) based on shorter times for Interval 2 Positioning—Incision (ORCT: 18.8, [18.1, 19.9] min, CBCT: 33.6, [32.2, 35.5] min; p < 0.0001) and Interval 4 Suture—Exit (ORCT: 24.3, [23.6, 26.1] min, CBCT: 29.3, [27.5, 30.7] min; p < 0.0001). Conclusions: The choice of imaging technology for navigated pedicle screw placement has significant impact on standard spine procedures even in a high-volume spine center with daily routine in such devices. Particularly with regard to the duration of surgeries, the shorter time needed for preparation and de-positioning in the ORCT group made the main difference, while the accuracy was even higher for the ORCT.Keywords
Funding Information
- Technische Universität München
This publication has 26 references indexed in Scilit:
- CT-navigation versus fluoroscopy-guided placement of pedicle screws at the thoracolumbar spine: single center experience of 4,500 screwsEuropean Spine Journal, 2012
- Pedicle screw navigation: a systematic review and meta-analysis of perforation risk for computer-navigated versus freehand insertionJournal of Neurosurgery: Spine, 2012
- Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniquesEuropean Spine Journal, 2011
- Intraoperative computed tomography image–guided navigation for posterior thoracolumbar spinal instrumentation in spinal deformity surgeryNeurosurgical Focus, 2010
- Intraoperative Computed Tomography With Integrated Navigation System in Spinal StabilizationsSpine, 2009
- INTRAOPERATIVE COMPUTED TOMOGRAPHY WITH INTEGRATED NAVIGATION SYSTEM IN A MULTIDISCIPLINARY OPERATING SUITEOperative Neurosurgery, 2009
- Cervical Pedicle Screws: Conventional Versus Computer-Assisted Placement of Cannulated ScrewsSpine, 2005
- Computer-assisted posterior instrumentation of the cervical and cervico-thoracic spineEuropean Spine Journal, 2004
- Accuracy of pedicle screw insertion with and without computer assistance: a randomised controlled clinical study in 100 consecutive patientsEuropean Spine Journal, 2000
- Accuracy of Pedicular Screw Placement In VivoSpine, 1990