Evaluation of Atlas-Based Attenuation Correction for Integrated PET/MR in Human Brain: Application of a Head Atlas and Comparison to True CT-Based Attenuation Correction
Open Access
- 22 October 2015
- journal article
- research article
- Published by Society of Nuclear Medicine in Journal of Nuclear Medicine
- Vol. 57 (2) , 215-220
Abstract
Introduction: Attenuation correction (AC) for integrated PET/MRI in human brain is still an open problem. In this study, we evaluated a simplified atlas-based AC (Atlas-AC) by comparing FDG PET data corrected using either Atlas-AC or true CT data (CT-AC). Methods: We enrolled 8 patients (median age 63). All patients underwent a clinically indicated whole body 18F-FDG-PET/CT (GE Healthcare Discovery 690 PET/CT) for staging, re-staging or follow-up of malignant disease. All patients volunteered for an additional PET/MR of the head (GE Healthcare SIGNA PET/MR). For each patient, 2 AC-maps were generated: an Atlas-AC map registered to a patient-specific LAVA-Flex MR sequence and using a vendor-provided head atlas generated from multiple CT head images, and a CT-based AC map. For comparative AC, the CT-AC map generated from PET/CT was superimposed on the Atlas-AC map. Using each AC map, PET images were reconstructed from the list-mode raw data from the PET/MRI scanner. All PET images were normalized to a brain template and FDG accumulation was quantified in 67 volumes-of-interest (VOIs; automated anatomical labeling atlas) using PMOD. Relative difference (%diff) between images based on Atlas-AC and CT-AC was calculated, and averaged difference images were generated. FDG uptake in all VOIs was compared using Bland-Altman analysis. Results: The range of error in all 536 VOIs was < 10% [range; -3.0% ~ 7.3%]. Whole brain FDG uptake based on Atlas-AC was slightly underestimated (%diff= 2.19±1.40%). The underestimation was most pronounced in the regions below the anterior/posterior commissure line, such as cerebellum, temporal lobe and central structures (%diff=3.69±1.43%, 3.25±1.42%, 3.05±1.18%). This suggests that Atlas-AC tends to underestimate the attenuation values of the skull base bone. Conclusion: The simplified Atlas-AC proved robust and was comparable to the gold-standard CT-AC, although there was a minor underestimation of tracer uptake values close to the base of the skull.Keywords
This publication has 31 references indexed in Scilit:
- Current status and future role of brain PET/MRI in clinical and research settingsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2015
- Simultaneous Whole-Body Time-of-Flight 18F-FDG PET/MRIClinical Nuclear Medicine, 2015
- Clinical applications of PET/MRI: current status and future perspectivesDiagnostic and Interventional Radiology, 2014
- Attenuation Correction Synthesis for Hybrid PET-MR Scanners: Application to Brain StudiesIEEE Transactions on Medical Imaging, 2014
- Combined PET/MR imaging in neurology: MR-based attenuation correction implies a strong spatial bias when ignoring boneNeuroImage, 2014
- MRI-Based Attenuation Correction for Hybrid PET/MRI Systems: A 4-Class Tissue Segmentation Technique Using a Combined Ultrashort-Echo-Time/Dixon MRI SequenceJournal of Nuclear Medicine, 2012
- Attenuation Correction Methods Suitable for Brain Imaging with a PET/MRI Scanner: A Comparison of Tissue Atlas and Template Attenuation Map ApproachesJournal of Nuclear Medicine, 2011
- MR‐based attenuation correction for hybrid PET‐MR brain imaging systems using deformable image registrationMedical Physics, 2010
- MRI-Based Attenuation Correction for PET/MRI: A Novel Approach Combining Pattern Recognition and Atlas RegistrationJournal of Nuclear Medicine, 2008
- Method for transforming CT images for attenuation correction in PET/CT imagingMedical Physics, 2006