Venovenous and venoarterial extracorporeal membrane oxygenation (ECMO) remains a crucial lifesaving therapy for critically ill neonates with severe cardiorespiratory failure. Both the roller pump as well as the centrifugal pump are safe and efficient systems, and some red blood cell breakdown and hemolysis occurs in all ECMO systems. The roller pump functions by gravity whereas the centrifugal pump promotes the flow of blood by a magnetically driven spinning rotor to generate negative pressure. Extracorporeal Life Support Organization data indicate a significant increase in intravascular hemolysis in neonatal and pediatric patients receiving ECMO when the centrifugal pump is used compared with its use in adults. Risk factors for developing hemolysis during ECMO are small cannula size, high negative inlet pressure in the pump head, and thrombosis in the pump head and oxygenator. Excessive red blood cell breakdown and release of plasma free hemoglobin (pfHb) saturate physiologic neutralizing mechanisms such as haptoglobin and hemopexin. The increase in pro-oxidant and proinflammatory pfHb levels causes endothelial dysfunction in a dose-dependent manner. Hemolysis also increases the risk of in-hospital morbidities such as renal injury, direct hyperbilirubinemia, and thrombosis without an increase in mortality in patients receiving ECMO. Hemolysis is an unavoidable side effect of current ECMO technology and there are no approved treatments or treatment guidelines for the neonatal population. Therefore, increased vigilance, recognition of the severity of the hemolytic process, and prompt management are essential to prevent severe endothelial injury leading to proinflammatory and prothrombotic events.Keywords
This publication has 29 references indexed in Scilit:
- Hemolysis in Pediatric Patients Receiving Centrifugal-Pump Extracorporeal Membrane OxygenationCritical Care Medicine, 2014
- Pediatric ECMO OutcomesASAIO Journal, 2013
- Outcomes of Neonates Undergoing Extracorporeal Membrane Oxygenation Support Using Centrifugal Versus Roller Blood PumpsThe Annals of Thoracic Surgery, 2012
- Extracorporeal life supportPediatric Critical Care Medicine, 2012
- Hemolytic and thrombocytopathic characteristics of extracorporeal membrane oxygenation systems at simulated flow rate for neonates*Pediatric Critical Care Medicine, 2012
- Hemolysis and ECMO pumps in the 21st CenturyPerfusion, 2010
- Haemoglobin blocks von Willebrand factor proteolysis by ADAMTS-13: A mechanism associated with sickle cell diseaseThrombosis and Haemostasis, 2009
- Hemolytic characteristics of three commercially available centrifugal blood pumpsPediatric Critical Care Medicine, 2005
- The Clinical Sequelae of Intravascular Hemolysis and Extracellular Plasma HemoglobinJAMA, 2005
- Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypassPerfusion, 2000