Long-length tomosynthesis and 3D-2D registration for intraoperative assessment of spine instrumentation
- 16 February 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (5) , 055008
Abstract
Purpose. A system for long-length intraoperative imaging is reported based on longitudinal motion of an O-arm gantry featuring a multi-slot collimator. We assess the utility of long-length tomosynthesis and the geometric accuracy of 3D image registration for surgical guidance and evaluation of long spinal constructs. Methods. A multi-slot collimator with tilted apertures was integrated into an O-arm system for long-length imaging. The multi-slot projective geometry leads to slight view disparity in both long-length projection images (referred to as ‘line scans’) and tomosynthesis ‘slot reconstructions’ produced using a weighted-backprojection method. The radiation dose for long-length imaging was measured, and the utility of long-length, intraoperative tomosynthesis was evaluated in phantom and cadaver studies. Leveraging the depth resolution provided by parallax views, an algorithm for 3D-2D registration of the patient and surgical devices was adapted for registration with line scans and slot reconstructions. Registration performance using single-plane or dual-plane long-length images was evaluated and compared to registration accuracy achieved using standard dual-plane radiographs. Results. Longitudinal coverage of ∼50–64 cm was achieved with a single long-length slot scan, providing a field-of-view (FOV) up to (40 × 64) cm2, depending on patient positioning. The dose-area product (reference point air kerma × x-ray field area) for a slot scan ranged from ∼702–1757 mGy·cm2, equivalent to ∼2.5 s of fluoroscopy and comparable to other long-length imaging systems. Long-length scanning produced high-resolution tomosynthesis reconstructions, covering ∼12–16 vertebral levels. 3D image registration using dual-plane slot reconstructions achieved median target registration error (TRE) of 1.2 mm and 0.6° in cadaver studies, outperforming registration to dual-plane line scans (TRE = 2.8 mm and 2.2°) and radiographs (TRE = 2.5 mm and 1.1°). 3D registration using single-plane slot reconstructions leveraged the ∼7–14° angular separation between slots to achieve median TRE ∼2 mm and Conclusion. The multi-slot configuration provided intraoperative visualization of long spine segments, facilitating target localization, assessment of global spinal alignment, and evaluation of long surgical constructs. 3D-2D registration to long-length tomosynthesis reconstructions yielded a promising means of guidance and verification with accuracy exceeding that of 3D-2D registration to conventional radiographs.Keywords
Funding Information
- Medtronic
This publication has 65 references indexed in Scilit:
- A review of 3D/2D registration methods for image-guided interventionsMedical Image Analysis, 2012
- Monitoring tumor motion by real time 2D/3D registration during radiotherapyRadiotherapy and Oncology, 2012
- Adult Spinal Deformity—Postoperative Standing ImbalanceSpine, 2010
- Parallax-free intra-operative X-ray image stitchingMedical Image Analysis, 2010
- Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patientsJournal of Neurosurgery: Spine, 2010
- The Prevalence of Wrong Level Surgery Among Spine SurgeonsSpine, 2008
- Prospective randomized comparison of radiation exposure from full spine radiographs obtained in three different techniquesEuropean Spine Journal, 2006
- Long Bone Panoramas From Fluoroscopic X-Ray ImagesIEEE Transactions on Medical Imaging, 2004
- Completely Derandomized Self-Adaptation in Evolution StrategiesEvolutionary Computation, 2001
- A comparison of similarity measures for use in 2-D-3-D medical image registrationIEEE Transactions on Medical Imaging, 1998