Flow and Leakage Characteristics of a Sashless Inclined Air-Curtain (sIAC) Fume Hood Containing Tall Pollutant-Generation Tanks
Authors:
- 25 August 2013
- journal article
- research article
- Published by Taylor & Francis in Journal of Occupational and Environmental Hygiene
- Vol. 10 (12) , 694-704
Abstract
In many fume hood applications, pollutant-generation devices are tall. Human operators of a fume hood must stand close to the front of the hood and lift up their hands to reach the top opening of the tall tank. In this situation, it is inconvenient to access the conventional hood because the sash acts as a barrier. Also, the bluff-body wake in front of the operator's chest causes a problem. By using laser-assisted smoke flow visualization and tracer-gas test methods, the present study examines a sashless inclined air-curtain (sIAC) fume hood for tall pollutant-generation tanks, with a mannequin standing in front of the hood face. The configuration of the sIAC fume hood, which had the important element of a backward-inclined push-pull air curtain, was different from conventional configurations. Depending on suction velocity, the backward-inclined air curtain had three characteristic modes: straight, concave, and attachment. A large recirculation bubble covering the area—from the hood ceiling to the work surface—was formed behind the inclined air curtain in the straight and concave modes. In the attachment mode, the inclined air curtain was attached to the rear wall of the hood, about 50 cm from the hood ceiling, and bifurcated into up and down streams. Releasing the pollutants at an altitude above where the inclined air curtain was attached caused the suction slot to directly draw up the pollutants. Releasing pollutants in the rear recirculation bubble created a risk of pollutants’ leaking from the hood face. The tracer-gas (SF6) test results showed that operating the sIAC hood in the attachment mode, with the pollutants being released high above the critical altitude, could guarantee almost no leakage, even though a mannequin was standing in front of the sashless hood face.Keywords
This publication has 19 references indexed in Scilit:
- Improving Flow and Spillage Characteristics of Range Hoods by Using an Inclined Air-Curtain TechniqueAnnals of Occupational Hygiene, 2010
- A Review of Published Quantitative Experimental Studies on Factors Affecting Laboratory Fume Hood PerformanceJournal of Occupational and Environmental Hygiene, 2008
- Aerodynamics and Performance Verifications of Test Methods for Laboratory Fume CupboardsAnnals of Occupational Hygiene, 2006
- Development and Evaluation of an Air-Curtain Fume Cabinet with Considerations of its AerodynamicsAnnals of Occupational Hygiene, 2006
- Required response time for variable air volume fume hood controllersAnnals of Occupational Hygiene, 2000
- Fume Hood Performance: Face Velocity Variability Inconsistent Air Volume SystemsApplied Occupational and Environmental Hygiene, 1998
- A REVIEW OF WAKE EFFECTS ON WORKER EXPOSUREAnnals of Work Exposures and Health, 1995
- CONTAINMENT TESTING OF FUME CUPBOARDS—II. TEST ROOM MEASUREMENTSAnnals of Work Exposures and Health, 1992
- CONTAINMENT TESTING OF FUME CUPBOARDS—I.METHODSAnnals of Work Exposures and Health, 1992
- A Study of Cold and Combusting Flow Around Bluff-Body CombustorsCombustion Science and Technology, 1987