The relationship between the functional loading rate and heel velocities was assessed in an active unilateral transfemoral amputee (UTFA) for adaptation to six different commercial prosthetic knees. To Investigate the short-term process of adaptability for UTFA for two types of prosthetic knees were evaluated, based on the correlation between heel vertical velocity and transient loading rate. The loading rate was calculated from the slope of ground reaction forces (GRF) and the corresponding time. The heel velocities and GRF were obtained by a motion analysis system. Biomechanical adaptation was evident following a short period of prosthetic knee use based upon the mean transient impact (loading rate) and the heel vertical velocity in slow, normal and fast walking. Trend lines of transient impact versus vertical heel velocity for a set of actively controlled variable damping (microprocessor) and mechanically passive prosthetic knees were all negatively correlated, except for an amputated leg during normal pace and healthy leg during fast pace. For an amputee to adapt well to a prescribed prosthesis excellent coordination between the intact and amputated limbs is required to control placement of the amputated leg to achieve a gait comparable to healthy subjects. There are many factors such as the hip, knee flexion/extension and the ankle plantarflexion/dorsiflexion contributing to the control of the transient impact of an amputee during walking. Therefore, for enhanced control of a prosthetic knee, a multifaceted approach is required. This study showed that UTFA adaption to different prosthetic knees in the short term with slower than self-selected speed is completely achievable based on the negative correlation of ground reaction forces versus linear velocity. Reduced speed may provide the prosthetists with the vision of the amputees' progression of adaptation with a newly prescribed prosthetic knee.Keywords
Funding Information
- University of Leeds
- Engineering and Physical Sciences Research Council (EP/K020463/1)
This publication has 29 references indexed in Scilit:
- Virtual prototyping of a semi-active transfemoral prosthetic legProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 2015
- Functional gait asymmetry of unilateral transfemoral amputeesHuman Movement Science, 2012
- Gait and balance of transfemoral amputees using passive mechanical and microprocessor-controlled prosthetic kneesGait & Posture, 2007
- Kinematic and kinetic comparisons of transfemoral amputee gait using C-Leg and Mauch SNS prosthetic kneesJournal of Rehabilitation Research and Development, 2006
- A Clinical Comparison of Variable-Damping and Mechanically Passive Prosthetic Knee DevicesAmerican Journal of Physical Medicine & Rehabilitation, 2005
- Psychosocial adjustment to lower-limb amputation: A reviewDisability and Rehabilitation, 2004
- Kinematic and Kinetic Variations of Below-Knee Amputee GaitJPO Journal of Prosthetics and Orthotics, 2002
- Mechanical work adaptations of above-knee amputee ambulationArchives of Physical Medicine and Rehabilitation, 1996
- Biomechanics of below-knee amputee gaitJournal of Biomechanics, 1988
- Joint forces in the human pelvis-leg skeleton during walkingJournal of Biomechanics, 1984