Determination of fluid leakages in the different screw-retained implant-abutment connections in a mechanical artificial mouth
- 15 July 2015
- journal article
- research article
- Published by Springer Nature in Journal of Materials Science: Materials in Medicine
- Vol. 26 (7) , 1-7
Abstract
This study shows the potential risk of microfiltration between two different types of implant-abutment connections screwed at 45 Ncm: external and internal. For the first time the use of a mechanical artificial mouth is used with the values (compression and torsion loads with a frequency of 2 Hz) of the human chewing. The mechanical tests were performed with an artificial saliva at 37 °C. The microgap in the connection was measured by an Image Analysis software incorporated in a high resolution scanning electron microscopy. Implant connections were filled with methylene blue by using self-adjustable precision pipettes and the quantity of leakage was determined by high sensitivity spectometry. We showed that the internal connection has lower microgaps compared to the external ones and these microgaps increased with the number of mechanical cycles. The leakage of methylene blue was higher when the external connection was performed. Microgaps and the influence of the mechanical loads are very important for the long-term behavior avoiding the bacteria colonization in the dental implants. These aspects should be known by the implantologists.Keywords
Funding Information
- Ministerio de Economía y Competitividad (MAT2012-30706)
This publication has 25 references indexed in Scilit:
- Analysis of the bacterial seal at the implant-abutment interface in external-hexagon and Morse taper-connection implants: an in vitro study using a new methodology.2012
- Leakage of saliva through the implant-abutment interface: in vitro evaluation of three different implant connections under unloaded and loaded conditions.2012
- Marginal Bone Level Changes at Dental Implants after 5 Years in Function: A Meta-AnalysisClinical Implant Dentistry and Related Research, 2011
- Bacterial Colonization of the Dental Implant Fixture–Abutment Interface: An In Vitro StudyJournal of Periodontology, 2009
- In vitro evaluation of the implant abutment connection sealing capability of different implant systemsJournal of Oral Rehabilitation, 2008
- Bone reactions to longstanding functional load at implants: an experimental study in dogsJournal of Clinical Periodontology, 2005
- Persistent Acute Inflammation at the Implant-Abutment InterfaceJournal of Dental Research, 2003
- Different types of inflammatory reactions in peri‐implant soft tissuesJournal of Clinical Periodontology, 1995
- Bacterial colonization of the internal part of two‐stage implants. An in vivo study.Clinical Oral Implants Research, 1993
- Osseointegrated Titanium Implants:Requirements for Ensuring a Long-Lasting, Direct Bone-to-Implant Anchorage in ManActa Orthopaedica, 1981