A High Dynamic Range 128 × 120 3-D Stacked CMOS SPAD Image Sensor SoC for Fluorescence Microendoscopy
- 1 March 2022
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Solid-State Circuits
- Vol. 57 (6) , 1649-1660
Abstract
A miniaturized 1.4 mm $\times \,\, 1.4$ mm, $128\times120$ single-photon avalanche diode (SPAD) image sensor with a five-wire interface is designed for time-resolved fluorescence microendoscopy. This is the first endoscopic chip-on-tip sensor capable of fluorescence lifetime imaging microscopy (FLIM). The sensor provides a novel, compact means to extend the photon counting dynamic range (DR) by partitioning the required bit depth between in-pixel counters and off-pixel noiseless frame summation. The sensor is implemented in STMicroelectronics 40-/90-nm 3-D-stacked backside-illuminated (BSI) CMOS process with 8- $\mu \text{m}$ pixels and 45% fill factor. The sensor capabilities are demonstrated through FLIM examples, including ex vivo human lung tissue, obtained at video rate.Keywords
Funding Information
- STMicroelectronics through the PROTEUS Project (EP/K03197X/1)
- Engineering and Physical Engineering Research Council (EP/S001638/1)
This publication has 30 references indexed in Scilit:
- Compact solid-state CMOS single-photon detector array for in vivo NIR fluorescence lifetime oncology measurementsBiomedical Optics Express, 2016
- Wireless fluorescence capsule for endoscopy using single photon-based detectionScientific Reports, 2015
- Octagonal CMOs image sensor with strobed RGB LED illumination for wireless capsule endoscopy2014
- Time-Domain Fluorescence Lifetime Imaging Techniques Suitable for Solid-State Imaging Sensor ArraysSensors, 2012
- Fluorescence Lifetime Techniques in Medical ApplicationsAnnals of Biomedical Engineering, 2012
- Swallowable Wireless Capsule Endoscopy: Progress and Technical ChallengesGastroenterology Research and Practice, 2011
- Upgrading time domain FLIM using an adaptive Monte Carlo data inflation algorithmCytometry Part A, 2011
- Miniaturized digital camera system for disposable endoscopic applicationsSensors and Actuators A: Physical, 2010
- A 2.6mW 2fps QVGA CMOS One-chip Wireless Camera with Digital Image Transmission Function for Capsule Endoscopes2006
- Solvent Dependence of the Fluorescence Lifetimes of Xanthene DyesPhotochemistry and Photobiology, 1999