Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy
- 27 May 2010
- journal article
- Published by Springer Nature in Journal of Neuro-Oncology
- Vol. 98 (2) , 277-285
Abstract
The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system’s on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.Keywords
This publication has 13 references indexed in Scilit:
- Endo-rectal balloon cavity dosimetry in a phantom: Performance under IMRT and helical tomotherapy beamsRadiotherapy and Oncology, 2009
- Quality Assurance of an Image Guided Intracranial Stereotactic Positioning SystemTechnology in Cancer Research & Treatment, 2009
- Dosimetric verification of helical tomotherapy for total scalp irradiationMedical Physics, 2008
- SU‐GG‐J‐75: Dosimetric and Image Quality Analysis of a New Ultrafine Imaging Mode in TomoTherapyMedical Physics, 2008
- Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy: An evaluation of the setup verification process for the special case of a rigid head phantomMedical Physics, 2006
- Intracranial stereotactic positioning systems: Report of the American Association of Physicists in Medicine Radiation Therapy Committee Task Group No. 68Medical Physics, 2005
- Commissioning and quality assurance of an optically guided three‐dimensional ultrasound target localization system for radiotherapyMedical Physics, 2002
- A technique for the quantitative evaluation of dose distributionsMedical Physics, 1998
- Linear Accelerator as a Neurosurgical Tool for Stereotactic RadiosurgeryNeurosurgery, 1988
- Stereotaxic Frame and Computer Software for CT-directed Neurosurgical LocalizationInvestigative Radiology, 1980