Evaluation of Physiological Amino Acids Profiling by Tandem Mass Spectrometry
Authors:
- 5 November 2013
- book chapter
- Published by Springer Nature
- Vol. 13, 119-128
Abstract
Background: Nowadays, the most conventional method to quantify physiological amino acids consists in ion exchange chromatography (IEC) followed by post-column ninhydrin derivatization and UV detection at two wavelengths. Unfortunately, the technique presents some drawbacks such as long run time, large sample volume, and specific costs associated to the maintenance of a dedicated instrument. Therefore, we aimed to switch towards a mass spectrometry approach. Methods: We have tested the aTRAQ kit for Amino Acid Analysis of Physiological Fluids (AB Sciex), affording the selective quantification of about 40 amino acids, and present here the results of our assessments. Results: Outlined accuracy profiles for each amino acid demonstrated very reliable data. A good linearity was observed from 1 to 1,000 μmol/L. Results comparison with IEC showed a right concordance. Reference intervals established were very similar to those obtained by IEC and patients suffering from inborn error of metabolism have been readily identified. Conclusions: The present approach offers a valid alternative to the IEC method, with several advantages: lower sample volume, run time reduction, and improved specificity. However, the aTRAQ method requires minute data reviewing, expending the overall time of procedure. Finally, financial and practical considerations of both techniques have to be counterbalanced before engaging any transition.Keywords
This publication has 17 references indexed in Scilit:
- Quantification of 22 plasma amino acids combining derivatization and ion-pair LC–MS/MSJournal of Chromatography B, 2011
- Rapid comprehensive amino acid analysis by liquid chromatography/tandem mass spectrometry: comparison to cation exchange with post‐column ninhydrin detectionRapid Communications in Mass Spectrometry, 2008
- Harmonization of strategies for the validation of quantitative analytical proceduresJournal of Pharmaceutical and Biomedical Analysis, 2007
- Harmonization of strategies for the validation of quantitative analytical proceduresJournal of Pharmaceutical and Biomedical Analysis, 2007
- Improvement of the decision efficiency of the accuracy profile by means of a desirability function for analytical methods validationAnalytica Chimica Acta, 2007
- Accuracy profiles from uncertainty measurementsTalanta, 2006
- Performances of a multidimensional on-line SPE-LC-ECD method for the determination of three major catecholamines in native human urine: Validation, risk and uncertainty assessmentsJournal of Chromatography B, 2006
- A new reversed‐phase liquid chromatographic/tandem mass spectrometric method for analysis of underivatised amino acids: evaluation for the diagnosis and the management of inherited disorders of amino acid metabolismRapid Communications in Mass Spectrometry, 2005
- Harmonization of strategies for the validation of quantitative analytical proceduresJournal of Pharmaceutical and Biomedical Analysis, 2004
- Assessment of tubular reabsorption of sodium, glucose, phosphate and amino acids based on spot urine samplesActa Paediatrica, 1994