Dynamic gating window technique for the reduction of dosimetric error in respiratory-gated spot-scanning particle therapy: An initial phantom study using patient tumor trajectory data
Authors:
Open Access
- 1 April 2020
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 21 (4) , 13-21
Abstract
Spot-scanning particle therapy possesses advantages, such as high conformity to the target and efficient energy utilization compared with those of the passive scattering irradiation technique. However, this irradiation technique is sensitive to target motion. In the current clinical situation, some motion management techniques, such as respiratory-gated irradiation, which uses an external or internal surrogate, have been clinically applied. In surrogate-based gating, the size of the gating window is fixed during the treatment in the current treatment system. In this study, we propose a dynamic gating window technique, which optimizes the size of gating window for each spot by considering a possible dosimetric error. The effectiveness of the dynamic gating window technique was evaluated by simulating irradiation using a moving target in a water phantom. In dosimetric characteristics comparison, the dynamic gating window technique exhibited better performance in all evaluation volumes with different effective depths compared with that of the fixed gate approach. The variation of dosimetric characteristics according to the target depth was small in dynamic gate compared to fixed gate. These results suggest that the dynamic gating window technique can maintain an acceptable dose distribution regardless of the target depth. The overall gating efficiency of the dynamic gate was approximately equal or greater than that of the fixed gating window. In dynamic gate, as the target depth becomes shallower, the gating efficiency will be reduced, although dosimetric characteristics will be maintained regardless of the target depth. The results of this study suggest that the proposed gating technique may potentially improve the dose distribution. However, additional evaluations should be undertaken in the future to determine clinical applicability by assuming the specifications of the treatment system and clinical situation.Keywords
Funding Information
- JSPS KAKENHI (17H04115)
This publication has 32 references indexed in Scilit:
- A serial 4DCT study to quantify range variations in charged particle radiotherapy of thoracic cancersJournal of Radiation Research, 2013
- Moving target irradiation with fast rescanning and gating in particle therapyMedical Physics, 2010
- A study on repainting strategies for treating moderately moving targets with proton pencil beam scanning at the new Gantry 2 at PSIPhysics in Medicine & Biology, 2010
- Gated Irradiation With Scanned Particle BeamsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Respiratory correlated cone-beam computed tomography on an isocentric C-armPhysics in Medicine & Biology, 2005
- Intensity modulated proton therapy: A clinical exampleMedical Physics, 2001
- An analytical approximation of the Bragg curve for therapeutic proton beamsMedical Physics, 1997
- Multiple Coulomb scattering of 160 MeV protonsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1993
- Effects of respiratory motion on dose uniformity with a charged particle scanning methodPhysics in Medicine & Biology, 1992