Intracavitary Cardiac Pacing

Authors:
  • NJNathan K. Jensen
    Nathan K. Jensen
  • WSW. Robot Schmidt
    W. Robot Schmidt
  • JGJoseph J. Garamella
    Joseph J. Garamella
  • MLMichael F. Lynch
    Michael F. Lynch
  • CPCharles A. Peterson
    Charles A. Peterson

Abstract
Complications developed in four patients with heart block managed by permanent transvenous cardiac pacemaking; the complications were penetration and perforation of the right ventricle, fracture of the catheter electrode, and pacemaker-induced ventricular tachycardia-fibrillation. Analysis of these problems led to recommendations for improvement in this method of pacing.