Intraoperative OCT Imaging of the Argus II Retinal Prosthesis System
- 1 November 2016
- journal article
- Published by SLACK, Inc. in Ophthalmic Surgery, Lasers and Imaging Retina
- Vol. 47 (11) , 999-1003
Abstract
BACKGROUND AND OBJECTIVE: Optimal placement of the Argus II Retinal Prosthesis System (Second Sight Medical Products, Sylmar, CA) is critical. Intraoperative optical coherence tomography (OCT) allows for intrasurgical visualization and confirmation of array placement. In this study, two different OCT systems were evaluated to assess the feasibility and utility of this technology during Argus II surgery. PATIENTS AND METHODS: Intraoperative OCT was performed on five patients undergoing Argus II implantation at Cole Eye Institute from June 2015 to July 2016. The EnVisu portable OCT (Bioptigen, Morrisville, NC) and microscope-integrated RESCAN 700 (Zeiss, Oberkochen, Germany) intraoperative OCT systems were utilized. The EnVisu was used in three patients and the RESCAN 700 in three of the five patients. Following array tacking, intraoperative OCT was performed over the entire array including the edges and tack. RESULTS: Intraoperative OCT allowed for visualization of the array/retina interface. Microscope integration of the OCT system facilitated ease of focusing, real-time feedback, surgeon-directed OCT scanning to the areas of interest, and enhanced image quality at points of interest. CONCLUSIONS: Intraoperative imaging of the Argus II electrode array is feasible and provides information about electrode array-retina interface and distance to help guide a surgeon. Microscope integration of OCT appears to provide an optimal and efficient approach to intraoperative OCT during Argus II array placement. [ Ophthalmic Surg Lasers Imaging Retina . 2016;47:999–1003.]Keywords
This publication has 8 references indexed in Scilit:
- Erratum to: Worldwide Argus II implantation: recommendations to optimize patient outcomesBMC Ophthalmology, 2016
- Ocular pulse amplitude and retina nerve fiber layer thickness in migraine patients without auraBMC Ophthalmology, 2016
- Ten-Year Follow-up of a Blind Patient Chronically Implanted with Epiretinal Prosthesis Argus IOphthalmology, 2015
- Argus II retinal prosthesis malrotation and repositioning with intraoperative optical coherence tomography in a posterior staphylomaClinical Ophthalmology, 2015
- Determination of Feasibility and Utility of Microscope-Integrated Optical Coherence Tomography During Ophthalmic SurgeryJAMA Ophthalmology, 2015
- Factors Affecting Perceptual Threshold in Argus II Retinal Prosthesis SubjectsTranslational Vision Science & Technology, 2013
- Retinal Prostheses: Current Clinical Results and Future NeedsOphthalmology, 2011
- Pathology of damaging electrical stimulation in the retinaExperimental Eye Research, 2007