An assay for neurotoxic esterase (neuropathy target esterase, NTE) was developed by Johnson (1,2) to assess the delayed neurotoxic potential of organophosphorus compounds. NTE activity is calculated from the rate of phenyl valerate hydrolysis resistant to paraoxon and sensitive to mipafox inhibition under specified conditions of inhibitor concentrations, pH, temperature, and incubation times with inhibitors and substrate. The amount of phenol produced is measured colorimetrically after its oxidative coupling with 4-aminoantipyrine to yield 4-N-(1,4-benzoquinoneimine)antipyrine, a chromophore with a wavelength of maximum absorbance (λm)510 nm and corresponding molar absorptivity (molar extinction coefficient, ɛ) equal to 13,900 M−1cm−1 The assay was improved and simplified later by Johnson (3) without any change in the λm or ɛ, even though the chromophore solvent was altered by adding the detergent, sodium dodecyl sulfate (SDS). The present work demonstrates that when the NTE assay is performed according to the improved procedure, with a final [SDS] of 3.0 mg/mL, the λm of the chromophore in the assay mixture is shifted from 510 to 490 nm. The same shift in the chromophore λm is observed when phenol standards are coupled with 4-aminoantipyrine in solutions containing 3.0 mg/mL SDS. A systematic investigation of the dependence of the λm of the chromophore on [SDS] in the assay mixture revealed that the spectral shift increases rapidly at an [SDS] greater than the apparent critical micelle concentration (CMC; estimated to be 0.53 mg/mL under these conditions) and begins to plateau at [SDS] greater than 10 mg/mL. Furthermore, the sensitivity of the assay was shown to be dependent upon [SDS], reaching a peak ɛ value of 16,470 M−1cm−1 at a λm of 486 nm when the final [SDS] was 9.5 mg/mL; it is therefore recommended that these optimized values of [SDS], λm, and ɛ be adopted in the NTE assay. Finally, because there are versions of assays for other enzymes or metabolites (e.g., acetylcholinesterase, butyrylcholinesterase, glucose, triglycerides, and uric acid) that rely on the same or a related chromophore as used in the NTE assay (4–8), the potential for detergent-induced shifts in the chromophore spectrum should be taken into account when optimizing these assays in the presence of detergents.