Impact of Different Pre‐Dilution Methods on Routine Blood Test Results: A Study With Sysmex XN ‐1000
Authors:
Open Access
- 2 June 2026
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- p. e70260
Abstract
Objectives: To compare three manual pre‐dilution procedures with the automated undiluted mode on the Sysmex XN‐1000 analyzer and to identify an optimal standard operating procedure for pre‐dilution testing. Methods: A total of 60 residual EDTA‐K3 whole‐blood specimens were consecutively collected between March and May 2024 and analyzed within 2 h. Each specimen was tested in the automated undiluted mode and by three manual pre‐dilution procedures: direct mixing, dry‐tip wiping, and wet‐tip wiping. Pre‐dilution was performed at 1:7 (20 μL of blood plus 120 μL of Sysmex DCL diluent) using calibrated pipettes. All procedures were performed by one trained technician. Each method was tested in triplicate, and mean values were analyzed. White blood cell count (WBC), red blood cell count (RBC), hemoglobin (HGB), platelet count (PLT), and hematocrit (HCT) were evaluated by paired t ‐test, Pearson correlation, Passing–Bablok regression, and bias analysis against CLIA allowable total error (TEa) criteria. Results: For WBC, RBC, HGB, and PLT, no significant difference was found between any pre‐dilution method and the undiluted mode (all p > 0.05); for HCT, only Method 1 differed significantly ( p = 0.022). Passing–Bablok regression showed that the wet‐tip wiping method had the best overall agreement with the undiluted mode. TEa‐based analysis showed that only Methods 2 and 3 met allowable bias limits for all five parameters. Conclusion: The wet‐tip wiping method showed the best overall analytical performance and may be preferred for low‐volume specimens on the Sysmex XN‐1000 platform.Keywords
This publication has 6 references indexed in Scilit:
- An Automated pre-Dilution Setup for Von Willebrand Factor Activity AssaysBio-Protocol, 2024
- Changes of platelet levels in routine blood tests may predict the possible remission and relapse of autoimmune bullous diseaseInternational Journal of Dermatology, 2022
- Routine Blood Tests in Asymptomatic Patients With Indolent Lymphoma Have Limited Ability to Detect Clinically Significant Disease ProgressionJCO Oncology Practice, 2020
- Improved accuracy in counting residual white blood cells in red cell concentrates using new blood bank mode software of SYSMEX XN‐1000 hematology analyzerTransfusion, 2020
- Routine blood tests are a basic screening tool for serious pathologyBMJ, 2017
- Verification and quality control of routine hematology analyzersInternational Journal of Laboratory Hematology, 2016