Test-Retest Reliability, Repeatability, and Sensitivity of an Automated Deformation-Controlled Indentation on Pressure Pain Threshold Measurement
Authors:
- Terry K. KooTKTerry K. Koo
- Jing-Yi GuoJGJing-Yi Guo
- Cameron M. BrownCBCameron M. Brown
- 1 February 2013
- journal article
- Published by National University of Health Sciences in Journal of Manipulative and Physiological Therapeutics
- Vol. 36 (2) , 84-90
Abstract
The purpose of this study was to construct a computerized deformation-controlled indentation system and compare its test-retest reliability, repeatability, and sensitivity with a manual algometer for pressure pain threshold (PPT) measurements. Pressure pain threshold measurements were made on 16 healthy subjects for 2 sessions on bilateral erector spinae muscles at L1, L3, and L5 spinal levels, consisting of 5 repeated trials each using computerized algometry on one side and manual algometry on the other side. Mean, SD, coefficient of variation, standard error of measurement, minimal detectable change, and intraclass correlation coefficient were calculated for both manual and computerized PPT measurements. Effects of session, level, method, and side on PPT measurements were evaluated using analysis of variance. Manual PPT measurements were significantly larger than computerized PPT measurements (P = .017), and session 2 was significantly larger than session 1 (P = .021). Coefficient of variation, intraclass correlation coefficient, standard error of measurement, and minimal detectable change of the manual and computerized PPT measurements were 10.3%, 0.91, 0.19 kg/cm(2), and 0.54 kg/cm(2) and 15.6%, 0.87, 0.26 kg/cm(2), and 0.73 kg/cm(2), respectively. Although computerized algometry offers the benefits of eliminating the effects of operator reaction time, operator anticipation, alignment error, and variation in indentation rate on PPT measurements, these results indicate that manual algometry using load-controlled strategy may be better than computerized deformation-controlled algometry in terms of test-retest reliability, repeatability, and sensitivity. Constant load-controlled indentation protocol may be more favorable for PPT measurements. Future computerized instrumentation for PPT measurements should adopt a load-controlled mechanism.Keywords
This publication has 32 references indexed in Scilit:
- Pressure thresholds of the human foot: measurement reliability and effects of stimulus characteristicsErgonomics, 2011
- Pressure pain sensitivity maps of the neck-shoulder and the low back regions in men and womenBMC Musculoskeletal Disorders, 2010
- Pain Perception Studies in Tension‐Type HeadacheHeadache: The Journal of Head and Face Pain, 2010
- Habituation to pain: Further support for a central componentPain, 2010
- Increased muscular and cutaneous pain sensitivity in cephalic region in patients with chronic tension‐type headacheEuropean Journal of Neurology, 2005
- Perception of surface pressure applied to the handErgonomics, 1999
- PRESSURE ALGOMETRY IN HEALTHY SUBJECTS: INTER-EXAMINER VARIABILITYJournal of Rehabilitation Medicine, 1998
- Reliability, Validity, Reproducibility of Algometry in Diagnosis of Active and Latent Tender Spots and Trigger PointsJournal of Musculoskeletal Pain, 1998
- Pressure pain thresholds in normal muscles: reliability, measurement effects, and topographic differencesPain, 1989
- Pain and pain control.Psychological Bulletin, 1977