Marker‐less real‐time intra‐operative camera and hand‐eye calibration procedure for surgical augmented reality
Open Access
- 12 November 2019
- journal article
- research article
- Published by Institution of Engineering and Technology (IET) in Healthcare Technology Letters
- Vol. 6 (6) , 255-260
Abstract
Accurate medical Augmented Reality (AR) rendering requires two calibrations, a camera intrinsic matrix estimation and a hand‐eye transformation. We present a unified, practical, marker‐less, real‐time system to estimate both these transformations during surgery. For camera calibration we perform calibrations at multiple distances from the endoscope, pre‐operatively, to parametrize the camera intrinsic matrix as a function of distance from the endoscope. Then, we retrieve the camera parameters intra‐operatively by estimating the distance of the surgical site from the endoscope in less than 1 s. Unlike in prior work, our method does not require the endoscope to be taken out of the patient; for the hand‐eye calibration, as opposed to conventional methods that require the identification of a marker, we make use of a rendered tool‐tip in 3D. As the surgeon moves the instrument and observes the offset between the actual and the rendered tool‐tip, they can select points of high visual error and manually bring the instrument tip to match the virtual rendered tool tip. To evaluate the hand‐eye calibration, 5 subjects carried out the hand‐eye calibration procedure on a da Vinci robot. Average Target Registration Error of approximately 7mm was achieved with just three data points.Keywords
This publication has 18 references indexed in Scilit:
- On mixed reality environments for minimally invasive therapy guidance: Systems architecture, successes and challenges in their implementation from laboratory to clinicComputerized Medical Imaging and Graphics, 2013
- Iterative Estimation of Rigid-Body TransformationsJournal of Mathematical Imaging and Vision, 2011
- A Robotic System for Intra-operative Trans-Rectal Ultrasound and Ultrasound Elastography in Radical Prostatectomy2011
- Hand-Eye Coordination Using a Video See-Through Augmented Reality SystemThe Ergonomics Open Journal, 2008
- Fully Automatic Endoscope Calibration for Intraoperative Use2006
- Laparoscope Self-calibration for Robotic Assisted Minimally Invasive Surgery2005
- A flexible new technique for camera calibrationIEEE Transactions on Pattern Analysis and Machine Intelligence, 2000
- Least-squares estimation of transformation parameters between two point patternsIEEE Transactions on Pattern Analysis and Machine Intelligence, 1991
- A new technique for fully autonomous and efficient 3D robotics hand/eye calibrationIEEE Transactions on Robotics and Automation, 1989
- Depth of focus, eye size and visual acuityVision Research, 1980