Predictors of Fracture Risk of a Small Caliber Implantable Cardioverter Defibrillator Lead
Authors:
- Arieh KerenAKArieh Keren
- Michael H. GollobMGMichael H. Gollob
- David H. BirnieDBDavid H. Birnie
- 1 April 2010
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 33 (4) , 437-443
Abstract
Introduction: The Sprint Fidelis 6949 implantable cardioverter defibrillator (ICD; Medtronic Inc., Minneapolis, MN, USA) lead has a high rate of fracture. Identification of predictors of subsequent fracture is useful in decision making about lead replacement and for future lead design. We sought to determine if there are clinical, procedural, or radiological features associated with a greater risk of subsequent lead fracture. Methods: Patients with Sprint Fidelis 6949 lead fractures (Fracture group) were identified from our institutional database. Each patient in the Fracture group was matched to two controls, immediately preceeding and succeeding Sprint Fidelis 6949 implant. Clinical and procedural characteristics were compared. Chest radiographs performed 2 weeks after ICD implant were reviewed by an observer blinded to outcomes. The following features were assessed: ICD tip location, lead slack, kinking of the lead body (≥90°), and presence of lead “crimping” within the anchoring sleeve. Results: Twenty-six patients with Sprint Fidelis 6949 lead fractures were identified and were matched to 52 control patients. On univariate analysis, a higher left ventricular ejection fraction (LVEF), prior ipsilateral device implant, history of prior ICD lead fracture, and noncephalic venous access were associated with risk of lead fracture. On multivariate analysis, a higher LVEF was the only independent predictor of lead fracture (P = 0.006). Radiological features were similar between the two groups. Conclusions: In this study, a higher LVEF was associated with a greater risk of lead fracture in patients with Sprint Fidelis 6949 ICD leads. Radiographic features did not predict subsequent risk of lead fracture in our population. (PACE 2010; 437–443)Keywords
This publication has 13 references indexed in Scilit:
- Increasing hazard of Sprint Fidelis implantable cardioverter-defibrillator lead failureHeart Rhythm, 2009
- Utilization of a national network for rapid response to the Medtronic Fidelis lead advisory: The Canadian Heart Rhythm Society Device Advisory CommitteeHeart Rhythm, 2009
- Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator LeadsCirculation, 2008
- Accelerating risk of Fidelis lead fractureHeart Rhythm, 2008
- Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory CommitteeHeart Rhythm, 2008
- To the EditorHeart Rhythm, 2008
- To the EditorHeart Rhythm, 2008
- Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator leadHeart Rhythm, 2007
- Ligature causing pseudofracture of a cardiac pacemaker lead.American Journal of Roentgenology, 1996
- Atrial “J” Pacing Lead Retention Wire Fracture: Radiographic Assessment, Incidence of Fracture, and Clinical ManagementPacing and Clinical Electrophysiology, 1995