Accuracy of Working Length Measurement Using Cone Beam Computed Tomography at Three Field of View Settings, Conventional Radiography, and Electronic Apex Locator: An Ex-vivo Study

Authors:

Abstract
Objective: Determining the working length (WL) in root canal treatment facilitates the treatment prognosis. The introduction of apex locators and new devices in dentistry influenced this consideration. This compar-ative study evaluated the accuracy of working length measurement by cone beam computed tomography (CBCT) in three fields of view (FOVs), conventional radiography, and the apex locator Raypex 5.Methods: The descriptive-analytical study was performed on 40 mandibular premolar teeth that met the inclusion criteria. Direct observation under the microscope was considered the gold standard and compared with measurements by the electronic apex locator, CBCT at three field of view settings and periapical images. The results were analyzed by paired t-tests and Wilcoxon tests. The level of significance was set at 0.05 (p=0.05).Results: Results showed that the CBCT at 5×5 FOV (p=0.733) and Conventional radiography(p=0.001) achieved the working lengths with the highest and lowest accuracy, respectively. In addition, the difference between actual and measured working length using conventional radiography was significant (p