Dual inversion recovery, ultrashort echo time (DIR UTE) imaging: Creating high contrast for short‐T2 species
Authors:
Open Access
- 23 January 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 63 (2) , 447-455
Abstract
Imaging of short‐T2 species requires not only a short echo time but also efficient suppression of long‐T2 species in order to maximize the short‐T2 contrast and dynamic range. This paper introduces a method of long‐T2 suppression using two long adiabatic inversion pulses. The first adiabatic inversion pulse inverts the magnetization of long‐T2 water and the second one inverts that of fat. Short‐T2 species experience a significant transverse relaxation during the long adiabatic inversion process and are minimally affected by the inversion pulses. Data acquisition with a short echo time of 8 μs starts following a time delay of inversion time (TI1) for the inverted water magnetization to reach a null point and a time delay of TI2 for the inverted fat magnetization to reach a null point. The suppression of long‐T2 species depends on proper combination of TI1, TI2, and pulse repetition time. It is insensitive to radiofrequency inhomogeneities because of the adiabatic inversion pulses. The feasibility of this dual inversion recovery ultrashort echo time technique was demonstrated on phantoms, cadaveric specimens, and healthy volunteers, using a clinical 3‐T scanner. High image contrast was achieved for the deep radial and calcified layers of articular cartilage, cortical bone, and the Achilles tendon. Magn Reson Med, 2010.Keywords
This publication has 24 references indexed in Scilit:
- Ultrashort TE spectroscopic imaging (UTESI): Application to the imaging of short T2 relaxation tissues in the musculoskeletal systemJournal of Magnetic Resonance Imaging, 2009
- Slice-selective tunable-flip adiabatic low peak-power excitation pulseMagnetic Resonance in Medicine, 2008
- Using adiabatic inversion pulses for long‐T2 suppression in ultrashort echo time (UTE) imagingMagnetic Resonance in Medicine, 2007
- In vivo MRI of submillisecond T2 species with two‐dimensional and three‐dimensional radial sequences and applications to the measurement of cortical bone waterNMR in Biomedicine, 2007
- Designing long‐T2 suppression pulses for ultrashort echo time imagingMagnetic Resonance in Medicine, 2006
- Three‐dimensional radial ultrashort echo‐time imaging with T2 adapted samplingMagnetic Resonance in Medicine, 2006
- T1, T2 relaxation and magnetization transfer in tissue at 3TMagnetic Resonance in Medicine, 2005
- MR Spectroscopic imaging of collagen: Tendons and knee menisciMagnetic Resonance in Medicine, 1995
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- Highly selective and π pulse generationJournal of Magnetic Resonance (1969), 1984