Axicon: a new laser beam delivery system for corneal surgery
Authors:
- 1 December 1990
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Quantum Electronics
- Vol. 26 (12) , 2305-2308
Abstract
A new laser beam delivery system using axicon beam shaping has been designed to produce controllable surface ablation as well as trephination for the corneal surgery. Noncontact cornea trephination was performed on human eye-bank eyes by using a Q-switched Er:YAG laser with a axicon-lens combination. The results shows that the uniformity of the trephination along the ring depends on the radial symmetry of the beam. The histological results of the samples show that the damage zone of the laser cut is less than 10 mu m. This demonstrates the possibility of replacing mechanical trephination in corneal surgery.<>Keywords
This publication has 18 references indexed in Scilit:
- Long-term Healing of the Central Cornea after Photorefractive Keratectomy Using an Excimer LaserOphthalmology, 1988
- Human Excimer Laser Keratectomy: Short-Term HistopathologyJournal of Refractive Surgery, 1988
- Excimer Laser Radial Keratotomy in the Living Human Eye: A Preliminary ReportJournal of Refractive Surgery, 1988
- Photorefractive keratectomy: A technique for laser refractive surgeryJournal of Cataract & Refractive Surgery, 1988
- Laser interactions with the corneaSurvey of Ophthalmology, 1986
- In vivo Experiments with the Excimer Laser – Technical Parameters and Healing ProcessesOphthalmologica, 1986
- Linear, annular, and radial focusing with axicons and applications to laser machiningApplied Optics, 1978
- Ring pattern of a lens–axicon doublet illuminated by a Gaussian beamApplied Optics, 1978
- Limbectomies, Keratectomies, and Keratostomies Performed with a Rapid-Pulsed Carbon Dioxide LaserAmerican Journal of Ophthalmology, 1971
- The Axicon: A New Type of Optical ElementJournal of the Optical Society of America, 1954