Abstract
Purpose: The measurement of corneal sensation can be useful in the diagnosis of corneal neuropathies. Although the Cochet–Bonnet (CB) esthesiometer is the gold standard, cotton wisp (CW) and dental floss are commonly used in clinical practice. Methods: We assessed the steady-state force of thick dental floss (tkDF), thin dental floss (tnDF), and CW compared with that of CB. We surveyed cornea/refractive specialists (CRS) and noncornea/refractive specialists (NCRS) for frequency and method by which corneal sensation was checked. Results: Compared with CB, the maximum force applied by tkDF was 0.92 grams more (P < 0.05), 0.15 grams more for tnDF (P < 0.05), and 0.23 grams less for CW (P < 0.05). Compared with CB, force per unit length was 0.13 g/mm higher for tkDF (P < 0.05), 0.001 g/mm higher for tnDF (P = 0.44), and 0.008 g/mm lower for CW (P = 0.10). We collected responses from 182/284 CRS and 102/284 NCRS. 12/182 CRS checked cornea sensation 0x/month compared with 42/102 NCRS. A total of 103/182 CRS and 54/102 NCRS checked cornea sensation 1-5x/month. CW was the most common tool used for CRS (115/170) and NCRS (53/60). Conclusions: tnDF provides the best analog for the CB. CW provides a small amount of force over a large range of lengths and may be a useful screening tool. tkDF is less suitable for screening. CRS check corneal sensation at a higher rate than NCRS. CW was the most common method used in both groups.