The work described here compared the available technical solutions for the treatment of ventricular tachycardia with stereotactic body radiation therapy. Due to the complexity of target motion during cardiac and pulmonary motion as well as the several proximate radio-sensitive structures of the tracheobronchial tree and esophagogastrointestinal tract, four potential candidates for this treatment were identified: Accuray CyberKnife (Accuray Incorporated, Sunnyvale, California, United States), Varian TrueBeam (Varian Medical Systems, Palo Alto, California, USA), Elekta Infinity (Elekta, Stockholm, Sweden), and Varian Edge (Varian Medical Systems, Palo Alto, California, USA). All four treatment modalities were evaluated for their ability to deliver a conformal, homogeneous dose to most of the target volume, to spare nearby and distant critical and sensitive anatomical structures as well as for treatment efficiency. It was found that conventional linear accelerator technology was superior in their ability to spare distant critical structures and deliver treatments efficiently while the CyberKnife showed superiority in sparing nearby critical structures more aggressively by creating larger dose gradients at the periphery of the target volume. Both treatment modalities were similar in their ability to cover the entire target with the prescription dose, conform that dose to the target volume, and deliver a homogeneous dose.Keywords
This publication has 11 references indexed in Scilit:
- Noninvasive Cardiac Radiation for Ablation of Ventricular TachycardiaThe New England journal of medicine, 2017
- Ablative Radiotherapy as a Noninvasive Alternative to Catheter Ablation for Cardiac ArrhythmiasCurrent Cardiology Reports, 2017
- The dosimetric comparison of Truebeam and Cyberknife treatment planning systems dose calculation accuracy for brain SRS treatment on RandophantomPhysica Medica, 2016
- Ventricular Tachycardia AblationJACC, 2016
- PrefaceJournal of the ICRU, 2014
- International Commission on Radiation Units and MeasurementsJournal of the ICRU, 2014
- Commissioning of the Varian TrueBeam linear accelerator: A multi‐institutional studyMedical Physics, 2013
- CyberKnife radiosurgery for an intracardiac metastasisBMJ Case Reports, 2011
- Accuray, Inc.: A Neurosurgical Business Case StudyCureus, 2009
- Ventricular Tachycardia and Sudden Cardiac DeathMayo Clinic Proceedings, 2009