Comparison of intra-aneurysmal flow modification using optical flow imaging to evaluate the performance of Evolve and Pipeline flow diverting stents
Authors:
Open Access
- 21 April 2020
- journal article
- research article
- Published by BMJ in Journal of NeuroInterventional Surgery
- Vol. 12 (8) , 814-817
Abstract
Background Flow diverting stent (FDS) devices have revolutionized the treatment of large and complex brain aneurysms, but there is still room for improvement, particularly on the flow diversion properties and technical challenges associated with stent deployment. In this study we compared flow diversion properties between the new generation Surpass Evolve (Stryker) and the Pipeline Flex (Medtronic) devices by quantitatively evaluating intra-aneurysmal flow modification. Methods An in vitro experimental set-up was used, consisting of four patient-specific silicone models with internal carotid aneurysms and a circulating hemodynamic simulation system with pulsatile flow. The Evolve and Pipeline stents were deployed across the neck of each aneurysm model, in a randomized fashion, for a total of eight device deployments. A 60 frames/s digital subtraction angiography run was acquired before and after placement of each FDS. An optical flow-analysis method was used to measure intra-aneurysmal flow modification induced by the stent by calculating a mean aneurysm flow amplitude (MAFA) before and after stent placement and computing a ratio. Results Average MAFA ratio values calculated from pre- and post-stent placement were significantly lower after deployment of the Evolve (n=4, mean=0.62±0.09) compared with the Pipeline device (n=4, mean=0.71±0.06) (p=0.03). Conclusions Our in vitro results show that the Evolve stent had a superior flow diversion effect compared with the Pipeline stent, which—based on clinical evidence—suggest it may promote faster aneurysm occlusion rates in patients.Keywords
This publication has 11 references indexed in Scilit:
- Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coveragePLOS ONE, 2018
- E-081 Assessment of the MAFA Ratio as a Quantitative Prognostic Marker of Aneurysm Occlusion after Flow Diverter Treatment2016
- Meta-Analysis of the Efficiency and Prognosis of Intracranial Aneurysm Treated with Flow Diverter DevicesJournal of Molecular Neuroscience, 2016
- Intra-aneurysmal hemodynamics: evaluation of pCONus and pCANvas bifurcation aneurysm devices using DSA optical flow imagingJournal of NeuroInterventional Surgery, 2015
- Metal coverage ratio of pipeline embolization device for treatment of unruptured aneurysms: Reality checkInterventional Neuroradiology, 2015
- New Generation of Flow Diverter (Surpass) for Unruptured Intracranial AneurysmsStroke, 2013
- Endovascular Treatment of Intracranial Aneurysms With Flow DivertersStroke, 2013
- A DSA-Based Method Using Contrast-Motion Estimation for the Assessment of the Intra-Aneurysmal Flow Changes Induced by Flow-Diverter StentsAmerican Journal of Neuroradiology, 2012
- Quantification of arterial flow using digital subtraction angiographyMedical Physics, 2012
- An Original Flow Diversion Device for the Treatment of Intracranial AneurysmsStroke, 2009