Impedance in the Diagnosis of Lead Malfunction
Authors:
- 1 February 2020
- journal article
- editorial
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 13 (2) , e008092
Abstract
Impedance is the ratio of voltage to current in an electrical circuit. Cardiovascular implantable electronic devices measure impedance to assess the structural integrity electrical performance of leads, typically using subthreshold pulses. We review determinants of impedance, how it is measured, variation in clinically measured pacing and high-voltage impedance and impedance trends as a diagnostic for lead failure and lead-device connection problems. We consider the differential diagnosis of abnormal impedance and the approach to the challenging problem of a single, abnormal impedance measurement. Present impedance provides a specific but insensitive diagnostic. For pacing circuits, we review the complementary roles of impedance and more sensitive oversensing diagnostics. Shock circuits lack a sensitive diagnostic. This deficiency is particularly important for insulation breaches, which may go undetected and present with short circuits during therapeutic shocks. We consider new methods for measuring impedance that may increase sensitivity for insulation breaches.Keywords
This publication has 41 references indexed in Scilit:
- Long-term electrical survival analysis of Riata and Riata ST silicone leads: National Veterans Affairs experienceHeart Rhythm, 2012
- Preventing Overdiagnosis of Implantable Cardioverter-Defibrillator Lead Fractures Using Device DiagnosticsJACC, 2011
- Late Perforation by Cardiac Implantable Electronic Device Leads: Clinical Presentation, Diagnostic Clues, and ManagementClinical Cardiology, 2010
- Prevalence and Characterization of Asymptomatic Pacemaker and ICD Lead Perforation on CTPacing and Clinical Electrophysiology, 2007
- High Pacing Impedances: Are You Overtorquing Your Leads?Pacing and Clinical Electrophysiology, 2005
- Implantable Cardioverter‐Defibrillator Malfunction due to Mechanical Failure of the Header ConnectionJournal of Cardiovascular Electrophysiology, 2004
- Detection and management of animplantable cardioverter defibrillator lead failureJACC, 2003
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996
- A New Pacemaker Algorithm for Continuous Capture Verification and Automatic Threshold Determination: Elimination of Pacemaker Afterpotential Utilizing a Triphasic Charge Balancing SystemPacing and Clinical Electrophysiology, 1992
- Air Entrapment as a Cause of Transient Cardiac Pacemaker MalfunctionPacing and Clinical Electrophysiology, 1979