Navigated MRI-guided liver biopsies in a closed-bore scanner: experience in 52 patients
Authors:
Open Access
- 12 November 2015
- journal article
- interventional
- Published by Springer Nature in European Radiology
- Vol. 26 (8) , 2462-2470
Abstract
Objectives To evaluate clinical effectiveness and diagnostic efficiency of a navigation device for MR-guided biopsies of focal liver lesions in a closed-bore scanner. Methods In 52 patients, 55 biopsies were performed. An add-on MR navigation system with optical instrument tracking was used for image guidance and biopsy device insertion outside the bore. Fast control imaging allowed visualization of the true needle position at any time. The biopsy workflow and procedure duration were recorded. Histological analysis and clinical course/outcome were used to calculate sensitivity, specificity and diagnostic accuracy. Results Fifty-four of 55 liver biopsies were performed successfully with the system. No major and four minor complications occurred. Mean tumour size was 23 ± 14 mm and the skin-to-target length ranged from 22 to 177 mm. In 39 cases, access path was double oblique. Sensitivity, specificity and diagnostic accuracy were 88 %, 100 % and 92 %, respectively. The mean procedure time was 51 ± 12 min, whereas the puncture itself lasted 16 ± 6 min. On average, four control scans were taken. Conclusions Using this navigation device, biopsies of poorly visible and difficult accessible liver lesions could be performed safely and reliably in a closed-bore MRI scanner. The system can be easily implemented in clinical routine workflow. Key Points • Targeted liver biopsies could be reliably performed in a closed-bore MRI. • The navigation system allows for image guidance outside of the scanner bore. • Assisted MRI-guided biopsies are helpful for focal lesions with a difficult access. • Successful integration of the method in clinical workflow was shown. • Subsequent system installation in an existing MRI environment is feasible.Keywords
This publication has 30 references indexed in Scilit:
- Magnetic resonance imaging of focal liver lesions: Approach to imaging diagnosisHepatology, 2010
- Flexible add‐on solution for MR image‐guided interventions in a closed‐bore scanner environmentMagnetic Resonance in Medicine, 2010
- Retrospektive Analyse der 5-Jahres-Ergebnisse von MRT-gesteuerten Biopsien im NiederfeldsystemRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2009
- Method for automatic localization of MR‐visible markers using morphological image processing and conventional pulse sequences: Feasibility for image‐guided proceduresJournal of Magnetic Resonance Imaging, 2007
- MRT-gezielte perkutane Biopsie bei unklaren fokalen Leberläsionen: Technik und ErgebnisseRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2003
- CT-gesteuerte perkutane Biopsien zur Klassifizierung von fokalen Leberläsionen: Vergleich zwischen 14 G- und 18 G-StanzbiopsienadelnRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1999
- Markierung nicht-palpabler Mammatumoren in der MRTRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1993
- Clinical use of a nonferromagnetic needle for magnetic resonance-guided biopsyGastrointestinal Radiology, 1989
- MR Body StereotaxisJournal of Computer Assisted Tomography, 1988
- A technique for MR-guided needle placementAmerican Journal of Roentgenology, 1988