Outcome of Polyester Cuff Retention Following Traction Removal of Tunneled Central Venous Catheters
Authors:
- Scott O. TrerotolaSTScott O. Trerotola
- Jan NamyslowskiJNJan Namyslowski
- Michael S. SteckerMSMichael S. Stecker
- Gordon McLennanGMGordon McLennan
- Nilesh H. PatelNPNilesh H. Patel
- Matthew S. JohnsonMJMatthew S. Johnson
- Himanshu ShahHSHimanshu Shah
- Roopa SeshadriRSRoopa Seshadri
- 1 June 2001
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 219 (3) , 651-654
Abstract
To elucidate the factors that contribute to cuff retention during traction removal of tunneled catheters, as well as to determine the risk of complication associated with polyester cuff retention. A total of 428 tunneled, cuffed catheters were removed with traction and local anesthesia. Polyester cuff retention was recorded when it occurred, and the effects of cuff retention were determined at a mean follow-up of 250 days. Statistical analysis was performed to determine the variables influencing cuff retention. Traction removal was successful in 428 (100%) patients. Of 428 catheters removed, 41 (10%) cuffs were retained. Silicone 10-F double-lumen and 9.6-F single-lumen catheters had a higher rate of cuff retention (27 [32%] of 84 and nine [39%] of 23, respectively) than did the split-tip polyurethane hemodialysis catheter (two [1%] of 196; P <.001). Cuff retention rates among other catheter types compared with that of the polyurethane catheter were not significantly different. Duration of catheter dwell did not significantly influence cuff retention. Of 41 retained cuffs, three required removal with cutdown for cuff migration to the exit site, which inhibited healing (n = 1); for suspected infection (n = 1); or for cosmetic purposes as requested by the patient (n = 1). The remaining patients had no complications associated with cuff retention. Traction removal of smaller-bore silicone catheters is more likely to result in cuff retention than removal of larger silicone and polyurethane catheters, and cuff retention is usually inconsequential.Keywords
This publication has 7 references indexed in Scilit:
- Mammography of breasts in which catheter cuffs have been retained: normal, infected, and postoperative appearances.American Journal of Roentgenology, 1997
- Outcome of tunneled hemodialysis catheters placed via the right internal jugular vein by interventional radiologists.Radiology, 1997
- Retained Hickman catheter cuff as a source of infectionJournal of Infection, 1993
- Mammographic appearance of the retained Dacron cuff of a Hickman catheter.American Journal of Roentgenology, 1990
- Hickman Catheter Complications in Marrow Transplant RecipientsJournal of Parenteral and Enteral Nutrition, 1986
- Complication of a Hickman CatheterJAMA, 1985
- The Hickman Indwelling CatheterAJN, American Journal of Nursing, 1980