Mechanical evaluation of a respiratory device
Authors:
- L.C. de LimaLLL.C. de Lima
- J.B.F. DuarteJDJ.B.F. Duarte
- F.P. Lépore NetoFNF.P. Lépore Neto
- P.T. AbePAP.T. Abe
- A.C. GastaldiAGA.C. Gastaldi
- 24 November 2004
- journal article
- research article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 27 (2) , 181-187
Abstract
This article aims to characterize the mechanical behaviour of the Flutter VRP1, a respiratory physiotherapy device designed to aid sputum clearance of the airways of patients. The device resembles a smoking pipe with a conical cavity containing a stainless steel sphere which floats up and down while the patient comes with a forced expiration through it. The sphere's oscillatory movement is function of the air flow rate and angular orientation of the device. When the sphere's oscillatory frequency matches the natural frequency of the patient's chestwall+abdomen system, it will produce resonance which, in turn, will enhance sputum clearance. A dynamical model of the Flutter was formulated and an experimental setup was assembled in order to study the oscillatory frequency of the sphere under different conditions of air flow rate, fluid pressure, angular orientation and sphere's material and weight. Interesting results presented by this article point to eventual mechanical optimization of the device and show information that could be beneficial to the professional of the respiratory physiotherapy.Keywords
This publication has 8 references indexed in Scilit:
- Chest physical therapy in patients with acute exacerbation of chronic bronchitis: Effectiveness of three methodsArchives of Physical Medicine and Rehabilitation, 2000
- The Intrapulmonary Percussive Ventilator and Flutter Device Compared to Standard Chest Physiotherapy in Patients wit Cystic FibrosisClinical Pediatrics, 1998
- Effect of High-Frequency Oral Airway and Chest Wall Oscillation and Conventional Chest Physical Therapy on Expectoration in Patients With Stable Cystic FibrosisChest, 1998
- Efficacy of the Flutter device for airway mucus clearance in patients with cystic fibrosisThe Journal of Pediatrics, 1994
- Use of Flutter VRP1 in the Management of Patients with Steroid-Dependent AsthmaRespiration, 1994
- The Flutter VRP1 Device for Post-thoracotomy PatientsPhysiotherapy, 1993
- Shock and Vibration Handbook (3rd ed.)Journal of Applied Mechanics, 1989
- Enhanced Tracheal Mucus Clearance with High Frequency Chest Wall CompressionAmerican Review of Respiratory Disease, 1983