Abstract
Phase wrapping is a crucial issue in Doppler optical coherence tomography (OCT) and restricts its automatic implementation for clinical applications that quantify total retinal blood flow. We propose an automated phase-unwrapping technique that takes advantage of the parabolic profile of blood flow velocity in vessels. Instead of inspecting the phase shift manually, the algorithm calculates the gradient magnitude of the phase shift on the cross-sectional image and automatically detects the presence of phase wrapping. The voxels affected by phase wrapping are corrected according to the determined flow direction adjacent to the vessel walls. We validated this technique in the rodent retina using a prototype visible-light OCT and in the human retina with a commercial infrared OCT system. We believe this signal processing method may well accelerate clinical applications of Doppler OCT in ophthalmology.Keywords
Funding Information
- Medical Center, University of Pittsburgh
- National Institutes of Health (DP3 DK104397)
- National Institutes of Health (DP3 DK104397)
- Oregon Health and Science Foundation (OHSU), National Institutes of Health (DP3 DK104397)
- National Eye Institute (DP3 DK104397)
- National Institutes of Health (R01 EY024544)
- National Institutes of Health (R01 EY024544)
- Oregon Health and Science Foundation (OHSU), National Institutes of Health (R01 EY024544)
- National Eye Institute (R01 EY024544)
- National Institutes of Health (P30 EY010572)
- National Institutes of Health (P30 EY010572)
- Oregon Health and Science Foundation (OHSU), National Institutes of Health (P30 EY010572)
- National Eye Institute (P30 EY010572)
- National Institutes of Health (R01 EY027833)
- National Eye Institute (R01 EY027833)
- National Institutes of Health (R01 EY010145)
- National Eye Institute (R01 EY010145)
This publication has 11 references indexed in Scilit:
- Fast and robust standard-deviation-based method for bulk motion compensation in phase-based functional OCTOptics Letters, 2018
- Angiographic and structural imaging using high axial resolution fiber-based visible-light OCTBiomedical Optics Express, 2017
- Two-dimensional phase unwrapping in Doppler Fourier domain optical coherence tomographyOptics Express, 2016
- En faceDoppler total retinal blood flow measurement with 70 kHz spectral optical coherence tomographyJournal of Biomedical Optics, 2015
- A combined method to quantify the retinal metabolic rate of oxygen using photoacoustic ophthalmoscopy and optical coherence tomographyScientific Reports, 2014
- Doppler Optical Coherence TomographyProgress in Retinal and Eye Research, 2014
- Efficient subpixel image registration algorithmsOptics Letters, 2008
- In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical Doppler tomographyOptics Express, 2003
- Real-time assessment of retinal blood flow with ultrafast acquisition by color Doppler Fourier domain optical coherence tomographyOptics Express, 2003
- Analysis of the phase unwrapping algorithmApplied Optics, 1982