A Fully Programmable Computing Architecture for Medical Ultrasound Machines
- 19 June 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Information Technology in Biomedicine
- Vol. 14 (2) , 538-540
Abstract
Application-specific ICs have been traditionally used to support the high computational and data rate requirements in medical ultrasound systems, particularly in receive beamforming. Utilizing the previously developed efficient front-end algorithms, in this paper, we present a simple programmable computing architecture, consisting of a field-programmable gate array (FPGA) and a digital signal processor (DSP), to support core ultrasound signal processing. It was found that 97.3% and 51.8% of the FPGA and DSP resources are, respectively, needed to support all the front-end and back-end processing for B-mode imaging with 64 channels and 120 scanlines per frame at 30 frames/s. These results indicate that this programmable architecture can meet the requirements of low- and medium-level ultrasound machines while providing a flexible platform for supporting the development and deployment of new algorithms and emerging clinical applications.Keywords
This publication has 10 references indexed in Scilit:
- New Demodulation Method for Efficient Phase-Rotation-Based BeamformingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2007
- New demodulation filter in digital phase rotation beamformingUltrasonics, 2006
- 6F-6 Thyroid Elastography Using Carotid Artery Pulsation: A Feasibility Study2006
- Compact FPGA-based beamformer using oversampled 1-bit A/D convertersIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2005
- Zone-based color flow imaging2004
- A single mediaprocessor-based programmable ultrasound systemIEEE Transactions on Information Technology in Biomedicine, 2003
- Ultrasound imaging systemThe Journal of the Acoustical Society of America, 2003
- Theoretical assessment of a synthetic aperture beamformer for real-time 3-D imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1999
- Blinded Comparison of an “Ultrasound Stethoscope” and Standard Echocardiographic InstrumentChest, 1988
- Real time ultrasonic imaging with a hand-held scanner Part I—Technical descriptionUltrasound in Medicine & Biology, 1978