Performance of a Radiation Protection Cabin During Implantation of Pacemakers or Cardioverter Defibrillators
- 26 March 2010
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 21 (4) , 428-430
Abstract
Radioprotection for Cardiac Device Implant. Introduction: Pacemaker implants are associated with a high cumulative exposure of the operators to radiation. Standard radiation protection with lead aprons is incomplete and the cause of spine disorders. A radiation protection cabin offers complete protection by surrounding the operator, without requiring a lead apron. Methods: We randomly and evenly assigned 60 patients undergoing implantations of permanent pacemakers or cardioverter defibrillators (ICD) with (a) a radiation protection cabin (cabin group, n = 30) versus (b) standard protection with a 0.5 mm lead‐equivalent apron (control group, n = 30). Radiation exposure was measured using personal electronic dosimeters placed on the thorax, back, and head of the operator. Results: The patient, procedural, and device characteristics of the 2 study groups were similar. All procedures in the cabin group were performed as planned without increase in duration or complication rate compared with the control group. The mean radiation dose to the head, normalized for fluoroscopy duration, was significantly lower in the cabin (0.040 ± 0.032 μSv/min) than in the control (1.138 ± 0.560 μSv/min) group (p < 0.0001). The radiation doses to the thorax (0.043 ± 0.027 vs 0.041 ± 0.040 μSv/min) and back (0.038 ± 0.029 vs 0.033 ± 0.018 μSv/min) in the cabin versus control group (under the apron) were similar. Conclusions: The use of a radiation protection cabin markedly decreased the exposure of the operator to radiation, and eliminated the need to wear a lead apron, without increasing the procedural time or complication rate during implantation of pacemaker and ICD. (J Cardiovasc Electrophysiol, Vol. 21, pp. 428–430, April 2010)Keywords
This publication has 12 references indexed in Scilit:
- Occupational Health Hazards in the Interventional Laboratory: Time for a Safer EnvironmentRadiology, 2009
- The non-cancer mortality experience of male workers at British Nuclear Fuels plc, 1946–2005International Journal of Epidemiology, 2008
- Medical exposure to ionising radiation and the risk of brain tumours: Interphone study group, GermanyEuropean Journal of Cancer, 2007
- Occupational exposure in the electrophysiology laboratory: quantifying and minimizing radiation burdenBritish Journal of Radiology, 2006
- Evaluation of a radiation protection cabin for invasive electrophysiological proceduresEuropean Heart Journal, 2006
- Increased Brain Cancer Risk in Physicians with High Radiation ExposureRadiology, 2005
- Somatic DNA damage in interventional cardiologists: a case‐control studyThe FASEB Journal, 2005
- Occupational hazards of interventional cardiologists: Prevalence of orthopedic health problems in contemporary practiceCatheterization and Cardiovascular Interventions, 2004
- Radiation exposure to medical staff in interventional and cardiac radiology.British Journal of Radiology, 1998
- Prevalence of Spinal Disc Disease Among Interventional CardiologistsThe American Journal of Cardiology, 1997