The variance of quantitative estimates in shear wave imaging: Theory and experiments
- 12 November 2012
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 59 (11) , 2390-2410
Abstract
In this paper, we investigate the relationship between the estimated shear modulus produced in shear wave imaging and the acquisition parameters. Using the framework of estimation theory and the Cramer-Rao lower bound applied both to the estimation of the velocity field variance and to the estimation of the shear wave travel time, we can derive the analytical formulation of the shear modulus variance σ μ 2 using relevant physical parameters such as the shear wave frequency, bandwidth, and ultrasonic parameters. This variance corresponds to the reproducibility of shear modulus reconstruction for a deterministic, quasi-homogeneous, and purely elastic medium. We thus consider the shear wave propagation as a deterministic process which is then corrupted during its observation by electronic noise and speckle decorrelation caused by shearing. A good correlation was found between analytical, numerical, and experimental results, which indicates that this formulation is well suited to understand the parameters' influence in those cases. The analytical formula stresses the importance of high-frequency and wideband shear waves for good estimation. Stiffer media are more difficult to assess reliably with identical acquisition signal-to-noise ratios, and a tradeoff between the reconstruction resolution of the shear modulus maps and the shear modulus variance is demonstrated. We then propose to use this formulation as a physical ground for a pixel-based quality measure that could be helpful for improving the reconstruction of real-time shear modulus maps for clinical applications.Keywords
This publication has 56 references indexed in Scilit:
- Shear Wave Elastography: A New Ultrasound Imaging Mode for the Differential Diagnosis of Benign and Malignant Thyroid NodulesThe Journal of Clinical Endocrinology & Metabolism, 2010
- Breast Lesions: Quantitative Elastography with Supersonic Shear Imaging—Preliminary ResultsRadiology, 2010
- Improving the Robustness of Time-of-Flight Based Shear Wave Speed Reconstruction Methods Using RANSAC in Human Liver in vivoUltrasound in Medicine & Biology, 2010
- A Kramers–Kronig-based quality factor for shear wave propagation in soft tissuePhysics in Medicine & Biology, 2009
- In vivo MR elastography of the prostate gland using a transurethral actuatorMagnetic Resonance in Medicine, 2009
- An error estimate on the direct inversion model in shear stiffness imagingInverse Problems, 2009
- Transient MR elastography (t‐MRE) using ultrasound radiation force: Theory, safety, and initial experiments in vitroMagnetic Resonance in Medicine, 2008
- Quantifying Hepatic Shear Modulus In Vivo Using Acoustic Radiation ForceUltrasound in Medicine & Biology, 2008
- Applications of magnetic resonance elastography to healthy and pathologic skeletal muscleJournal of Magnetic Resonance Imaging, 2007
- In Vivo Determination of Hepatic Stiffness Using Steady-State Free Precession Magnetic Resonance ElastographyInvestigative Radiology, 2006