An investigation of beam parameters for Co-60 tomotherapy
Author:
- 17 September 2007
- journal article
- Published by Wiley in Medical Physics
- Vol. 34 (10) , 3838-3843
Abstract
The effect of various physical beam parameters for Co-60 tomotherapy is examined, including: Cylindrical source size, source-to-collimator distance (SCD) and collimator leaf width. In general, beam profile effects are seen with larger Co-60 sources that are not seen with conventional linacs and multileaf collimators that are used for IMRT, including a broadening of the profile width as the source width increases. A treatment planning study was conducted to evaluate the effect of various beam parameter combinations with planning regions of interest typical of a simultaneous boost head and neck treatment. Combinations of SCD and source FWHM that produce an 80%-20% profile distance less than approximately 0.85 cm at 0.5 cm depth produced a Co-60 tomotherapy plan with better DVH results than a conventional 7-field linac plan for a 1 cm leaf width. Further improvement may be achieved by reducing the leaf width. In general it may be concluded that commercially available cylindrical Co-60 sources of 1.5-2.0 cm diameter may be appropriate for tomotherapy if the proper combination of beam parameters is chosen.Keywords
This publication has 8 references indexed in Scilit:
- Intensity-modulated radiation therapy (IMRT) dosimetry of the head and neck: A comparison of treatment plans using linear accelerator–based IMRT and helical tomotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- SU-FF-T-336: Multi-Slice Tomotherapy Delivery Methods Using a Novel Multileaf CollimatorMedical Physics, 2006
- SU-DD-A1-01: Advances in Co-60 Based Tomotherapy Including Megavoltage CTMedical Physics, 2006
- Significant improvement in normal tissue sparing and target coverage for head and neck cancer by means of helical tomotherapyRadiotherapy and Oncology, 2006
- Magnetic resonance imaging for adaptive cobalt tomotherapy: A proposalJournal of Medical Physics, 2006
- Comparing step-and-shoot IMRT with dynamic helical tomotherapy IMRT plans for head-and-neck cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Radiation characteristics of helical tomotherapyMedical Physics, 2004
- Tomotherapy: A new concept for the delivery of dynamic conformal radiotherapyMedical Physics, 1993