Feasibility of absolute quantification for 31P MRS at 7 T
Open Access
- 20 March 2019
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 82 (1) , 49-61
Abstract
Purpose: Phosphorus spectroscopy can differentiate among liver disease stages and types. To quantify absolute concentrations of phosphorus metabolites, sensitivity calibration and transmit field () correction are required. The trend toward ultrahigh fields (7 T) and the use of multichannel RF coils makes this ever more challenging. We investigated the constraints on reference phantoms, and implemented techniques for the absolute quantification of human liver phosphorus spectra acquired using a 10‐cm loop and a 16‐channel array at 7 T.Methods: The effect of phantom conductivity was assessed at 25.8 MHz (1.5 T), 49.9 MHz (3 T), and 120.3 MHz (7 T) by electromagnetic modeling. Radiofrequency field maps () were measured in phosphate phantoms (18 mM and 40 mM) at 7 T. These maps were used to assess the correction of 4 phantom 3D‐CSI data sets using 3 techniques: phantom replacement, explicit normalization, and simplified normalization. In vivo liver spectra acquired with a 10‐cm loop were corrected with all 3 methods. Simplified normalization was applied to in vivo 16‐channel array data sets.Results: Simulations show that quantification errors of less than 3% are achievable using a uniform electrolyte phantom with a conductivity of 0.23‐0.86 S.m−1 at 1.5 T, 0.39‐0.58 S.m−1 at 3 T, and 0.34‐0.42 S.m−1 (16‐19 mM KH2PO4(aq)) at 7 T. The mean γ‐ATP concentration quantified in vivo at 7 T was 1.39 ± 0.30 mmol.L−1 to 1.71 ± 0.35 mmol.L−1 wet tissue for the 10‐cm loop and 1.88 ± 0.25 mmol.L−1 wet tissue for the array.Conclusion: It is essential to select a calibration phantom with appropriate conductivity for quantitative phosphorus spectroscopy at 7 T. Using an 18‐mM phosphate phantom and simplified normalization, human liver phosphate metabolite concentrations were successfully quantified at 7 T.Keywords
Funding Information
- Wellcome Trust (098436, 12, 2, 0001, 17, 15-0029)
- Royal Society
- Medical Research Council
This publication has 38 references indexed in Scilit:
- Human cardiac 31P magnetic resonance spectroscopy at 7 teslaMagnetic Resonance in Medicine, 2013
- Receive array magnetic resonance spectroscopy: Whitened singular value decomposition (WSVD) gives optimal Bayesian solutionMagnetic Resonance in Medicine, 2010
- Interactions and dynamics in electrolyte solutions by dielectric spectroscopyPhysical Chemistry Chemical Physics, 2009
- Dielectric Spectroscopy of Aqueous Solutions of KCl and CsClThe Journal of Physical Chemistry A, 2003
- Quantification of phosphorus metabolites from chemical shift imaging spectra with corrections for point spread effects and B1 inhomogeneityMagnetic Resonance in Medicine, 1998
- The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissuesPhysics in Medicine & Biology, 1996
- Molar Quantitation of Hepatic MetabolitesIn Vivo in Proton-decoupled, Nuclear Overhauser Effect Enhanced31P NMR Spectra Localized by Three-dimensional Chemical Shift ImagingNMR in Biomedicine, 1996
- A 31P and 1H-NMR investigation in vitro of normal and abnormal human liverBiochimica et Biophysica Acta (BBA) - Molecular Basis of Disease, 1993
- Quantitative P‐31 MR spectroscopy of the liver in alcoholic cirrhosisJournal of Magnetic Resonance Imaging, 1992
- Non‐invasive quantitation of human liver metabolites using image‐guided 31p magnetic resonance spectroscopyNMR in Biomedicine, 1990