Development and signals simulation of a diamond detector neutral particle analyzer on Huanliu-2A (HL-2A) tokamak
- 1 November 2022
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 93 (11) , 113507
Abstract
A new neutral particle analyzer (NPA) diagnostic based on single crystal chemical vapor deposition (sCVD) diamond detector that provides measurements of fast ions has been designed and installed on HL-2A tokamak. Diamond detectors have been applied in some magnetic confinement fusion devices due to their outstanding properties of compact size and radiation hardness. This DNPA can measure energies above 13.4 keV. The line of sight (LOS) of the DNPA intersects with the NBI No. 2 with a tangency radius of 154.8 cm. Due to the pitch angle defined by the LOS and geometry of the diagnostic, the DNPA is mainly sensitive to trapped ions. To interpret the energy spectrum and verify the feasibility of the design of the DNPA, a Monte Carlo code called FIDASIM, which is a synthetic diagnostic code that simulates fast ion D-alpha and NPA signals, is applied to model the neutral flux reaching the detector. The results show that the flux is mainly contributed by the low energy fast ions (E < 10 keV) and it is mainly coming from the active components, the passive signal is dominant in the high energy region (E > 15 keV). The modeling features the ability to distinguish between active and passive signals, and the simulated strong passive signals are suggested to come from charge exchange between cold neutrals and fast ions around the plasma edge. In addition, despite the large ratio of halo neutrals, essentially it has a limited contribution to the energy spectrum.Keywords
Funding Information
- National MCF Energy R&D Program (2019YFE03020003)
This publication has 23 references indexed in Scilit:
- Fast-ion Dα measurements and simulations in quiet plasmasPhysics of Plasmas, 2007
- Overview of HL-2A experiment resultsNuclear Fusion, 2007
- Solid state neutral particle analyzer array on National Spherical Torus ExperimentReview of Scientific Instruments, 2004
- SRIM-2003Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2004
- The tokamak Monte Carlo fast ion module NUBEAM in the National Transport Code Collaboration libraryComputer Physics Communications, 2004
- Application of natural diamond detector to energetic neutral particle measurements on NSTXReview of Scientific Instruments, 2003
- Tokamak Fusion Test Reactor charge exchange atom spectrometry using a natural diamond detectorReview of Scientific Instruments, 1999
- The behaviour of fast ions in tokamak experimentsNuclear Fusion, 1995
- A Monte Carlo computer program for the transport of energetic ions in amorphous targetsNuclear Instruments and Methods, 1980
- A neutral-particle analyser for plasma diagnosticsJournal of Physics E: Scientific Instruments, 1978