Independent quality assurance of a helical tomotherapy machine using the dose magnifying glass
- 30 March 2011
- journal article
- Published by Wiley in Medical Physics
- Vol. 38 (4) , 2256-2264
Abstract
Purpose:: Helical tomotherapy is a complex delivery technique, integrating CT image guidance and intensity modulated radiotherapy in a single system. The integration of the CT detector ring on the gantry not only allows patient position verification but is also often used to perform various QA procedures. This convenience lacks the rigor of a machine‐independent QA process.Methods:: In this article, a Si strip detector, known as the Dose Magnifying Glass (DMG), was used to perform machine‐independent QA measurements of the multileaf collimator alignment, leaf open time threshold, and leaf fluence output factor (LFOF).Results:: The DMG measurements showed good agreements with EDR2 film for the MLC alignment test while the CT detector agrees well with DMG measurements for leaf open time threshold and LFOF measurements. The leaf open time threshold was found to be approximately 20 ms. The LFOF measured with the DMG agreed within error with the CT detector measured LFOF.Conclusions:: The DMG with its 0.2 mm spatial resolution coupled to TERA ASIC allowed real‐time high temporal resolution measurements of the tomotherapy leaf movement. In conclusion, DMG was shown to be a suitable tool for machine‐independent QA of a tomotherapy unit.Keywords
Funding Information
- Australian Institute of Nuclear Science and Engineering (AINGRA08039)
This publication has 12 references indexed in Scilit:
- QA for helical tomotherapy: Report of the AAPM Task Group 148a)Medical Physics, 2010
- A silicon strip detector dose magnifying glass for IMRT dosimetryMedical Physics, 2010
- Helical Tomotherapy Quality AssuranceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Helical tomotherapy dynamic quality assuranceMedical Physics, 2006
- History of tomotherapyPhysics in Medicine & Biology, 2006
- Quality assurance of a helical tomotherapy machinePhysics in Medicine & Biology, 2004
- Radiation characteristics of helical tomotherapyMedical Physics, 2004
- Clinical helical tomotherapy commissioning dosimetryMedical Physics, 2003
- Benchmarking beam alignment for a clinical helical tomotherapy deviceMedical Physics, 2003
- On the verification of the incident energy fluence in tomotherapy IMRTPhysics in Medicine & Biology, 2001