Incidence of Focal Osteolysis in a Titanium Limb-Lengthening Device: A Comparison to a Stainless-Steel Model
- 1 December 2023
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 37 (12) , 664-668
Abstract
Objectives: To evaluate 1 manufacturer's titanium intramedullary lengthening nail (ILN) for radiographic evidence of adverse tissue reactions similar to those in their stainless-steel ILN that was recalled in 2021, with specific outcomes of (1) incidence of periosteal reaction and osteolysis and (2) adverse tissue reaction-related device failures. Design: Radiographic review of consecutive patients with ≥1-year clinical follow-up who received 1 manufacturer's ILNs implanted in the femur and/or tibia from February 2012 to July 2021. Setting: Single tertiary-care center. Patients/Participants: Two hundred twenty-six limbs with titanium ILNs and 57 limbs with stainless-steel ILNs. Intervention: Precice titanium ILN (NuVasive Specialized Orthopedics, San Diego, CA) and Stryde stainless-steel ILN (NuVasive Specialized Orthopedics). Main Outcome Measurements: Radiographic outcome data and complications. Results: Of 226 limbs with titanium ILNs, 160 (70.8%) had periosteal reaction, and 2 (0.9%) had osteolysis. Periosteal reaction and osteolysis rates in stainless-steel ILNs (57 limbs) were 39% and 26%, respectively. No adverse tissue reaction-related device failures were noted with titanium. Several other complications presented, notably hip and/or knee tissue contracture (141), consolidation problems (8), and needing peroneal nerve decompression (14). Conclusions: Osteolysis incidence was substantially less among the titanium population, supporting the claim that observed osteolysis in the stainless-steel cohort was attributed to its construct. On average, periosteal reaction was radiographically appreciated later in patients with titanium than stainless steel. Further study is needed to confirm this finding. Level of Evidence: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.Keywords
This publication has 25 references indexed in Scilit:
- Load-bearing in cortical bone microstructure: Selective stiffening and heterogeneous strain distribution at the lamellar levelJournal of the Mechanical Behavior of Biomedical Materials, 2013
- Limb Lengthening and Then Insertion of an Intramedullary Nail: A Case-matched ComparisonClinical Orthopaedics and Related Research, 2008
- Histomorphometric analysis of the intramedullary bone response to titanium particles in wild‐type and IL‐1R1 knock‐out mice: A preliminary studyJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2007
- The role of macrophages in osteolysis of total joint replacementBiomaterials, 2005
- A literature review of the association between wear rate and osteolysis in total hip arthroplastyThe Journal of Arthroplasty, 2002
- Metal Versus Polyethylene Wear Particles in Total Hip ReplacementsClinical Orthopaedics and Related Research, 1996
- Tissue response in relation to type of wear particles around failed hip arthroplastiesThe Journal of Arthroplasty, 1990
- Problems, Obstacles, and Complications of Limb Lengthening by the Ilizarov TechniqueClinical Orthopaedics and Related Research, 1990
- The Tension-Stress Effect on the Genesis and Growth of TissuesClinical Orthopaedics and Related Research, 1989