Relationship between plasma and salivary melatonin and cortisol investigated by LC-MS/MS
Authors:
- Martijn van FaassenMFMartijn van Faassen
- Rainer BischoffRBRainer Bischoff
- Ido P. KemaIKIdo P. Kema
- 20 December 2016
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 55 (9) , 1340-1348
Abstract
Background: Disturbance of the circadian rhythm has been associated with disease states, such as metabolic disorders, depression and cancer. Quantification of the circadian markers such as melatonin and cortisol critically depend on reliable and reproducible analytical methods. Previously, melatonin and cortisol were primarily analyzed separately, mainly using immunoassays.Methods: Here we describe the validation and application of a high-throughput liquid chromatography in combination with mass spectrometry (LC-MS/MS) method for the combined analysis of melatonin and cortisol in plasma and saliva. The LC-MS/MS method was validated according to international validation guidelines. We used this method to analyze total plasma, free plasma (as obtained by equilibrium dialysis) and saliva melatonin and cortisol in healthy adults.Results: Validation results for plasma and saliva melatonin and cortisol were well within the international validation criteria. We observed no difference between saliva collected by passive drooling or Salivette. Moreover, we noted a significant difference in saliva vs. free plasma melatonin. We observed on average 36% (95% CI: 4%–60%) higher salivary melatonin levels in comparison to free plasma melatonin, suggestive of local production of melatonin in the salivary glands.Conclusions: The novel outcome of this study is probably due to the high precision of our LC-MS/MS assay. These outcomes illustrate the added value of accurate and sensitive mass spectrometry based methods for the quantification of neuroendocrine biomarkers.Keywords
This publication has 47 references indexed in Scilit:
- Quantitation of melatonin and n‐acetylserotonin in human plasma by nanoflow LC‐MS/MS and electrospray LC‐MS/MSJournal of Mass Spectrometry, 2012
- Accuracy of 6 Routine 25-Hydroxyvitamin D Assays: Influence of Vitamin D Binding Protein ConcentrationClinical Chemistry, 2012
- Simultaneous analysis of cortisol and cortisone in saliva using XLC–MS/MS for fully automated online solid phase extractionJournal of Chromatography B, 2012
- An ultra-high sensitive bioanalytical method for plasma melatonin by liquid chromatography–tandem mass spectrometry using water as calibration matrixJournal of Chromatography B, 2011
- Effects of saliva collection using cotton swabs on melatonin enzyme immunoassayJournal of Circadian Rhythms, 2011
- Salivary Cortisone Is a Potential Biomarker for Serum Free CortisolThe Journal of Clinical Endocrinology & Metabolism, 2010
- Evaluation of saliva collection devices for the analysis of steroids, peptides and therapeutic drugsJournal of Pharmaceutical and Biomedical Analysis, 2008
- Free Cortisol Levels after Awakening: A Reliable Biological Marker for the Assessment of Adrenocortical ActivityLife Sciences, 1997
- A General Regression Procedure for Method Transformation. Application of Linear Regression Procedures for Method Comparison Studies in Clinical Chemistry, Part IIIClinical Chemistry and Laboratory Medicine (CCLM), 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986