A Minimal Model of the Monophasic Defibrillation Pulse
Author:
- 1 April 1993
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (4) , 769-777
Abstract
A minimal model of the defibrillation capability of a monophasic capacitive discharge pulse is derived from the Weiss‐Lapicque strength duration model. The model suggests that present, empirically derived values of pulse durations and tilts are close to optimum for presently used values of capacitors and electrode resistances. The model suggests that neither the tilt nor fixed duration specification is universally superior to the other for dealing with electrode resistance changes. A tilt specification would appear to best handle resistance decreases while a fixed duration specification would best handle resistance increases. The model was used to study the effect of capacitance changes. It appears that the optimum tilt and pulse duration vary with the capacitance value. The model further suggests that decreasing the capacitance from presently used values may lower defibrillation thresholds.Keywords
This publication has 17 references indexed in Scilit:
- Optical recordings in the rabbit heart show that defibrillation strength shocks prolong the duration of depolarization and the refractory period.Circulation Research, 1991
- Strength-duration and probability of success curves for defibrillation with biphasic waveforms.Circulation, 1990
- Mechanism of cardiac defibrillation in open-chest dogs with unipolar DC-coupled simultaneous activation and shock potential recordings.Circulation, 1990
- Ventricular defibrillation using biphasic waveforms: The importance of phasic durationJACC, 1989
- Double and triple sequential shocks reduce ventricular defibrillation threshold in dogs with and without myocardial infarctionJACC, 1986
- Comparison of the defibrillation threshold and the upper limit of ventricular vulnerability.Circulation, 1986
- Internal cardiac defibrillation in man: pronounced improvement with sequential pulse delivery to two different lead orientations.Circulation, 1986
- Efficacy and safety of defibrillation with rectangular waves of 2− to 20-milliseconds durationCritical Care Medicine, 1983
- Transthoracic ventricular defibrillation in the 100 kg calf with unidirectional rectangular pulses.Circulation, 1977
- Termination of ventricular fibrillation in dogs by depolarizing a critical amount of myocardiumThe American Journal of Cardiology, 1975