Midwave-infrared snapshot imaging spectrometer
Authors:
- 1 September 2001
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 40 (25) , 4501-4506
Abstract
We report results from a demonstration of a midwave-infrared, nonscanning, high-speed imaging spectrometer capable of simultaneously recording spatial and spectral data from a rapidly varying target scene. We demonstrated high-speed spectral imaging by collecting spectral and spatial snapshots of blackbody targets and combustion products. The instrument is based on computed tomography concepts and operates in a midwave-infrared band of 3.0–5.0 µm. We record raw images at a frame rate of 60 frames/s, using a 512 × 512 InSb focal-plane array. Reconstructed object cube estimates were sampled at 46 × 46 × 21 (x, y, λ) elements, or 0.1-µm spectral sampling. Reconstructions of several objects are presented.Keywords